Abstract

OBJECTIVE

To assess the effect of reduced contrast sensitivity on three commonly used colour vision tests in order to establish key discrepancies that may be relevant for clinical practice.

METHODS

A prospective non-interventional clinical study of colour vision testing using three commonly used devices: Ishihara and Hardy-Rand-Rittler (H-R-R) pseudoisocochromatic plate tests, and Farnsworth D-15 arrangement test performed under progressively reduced contrast sensitivity conditions achieved with a neutral density filter bar.

RESULTS

The Pelli-Robson contrast sensitivity (PRCS) at which 5% of the population should first experience a 10% reduction in colour vision testing from baseline was calculated for each of the three colour vision devices: Farnsworth D-15 test: 1.81 log contrast sensitivity (CS), H-R-R test: 1.69 log CS, and Ishihara test: 1.34 log CS. Single factor repeated measures analyses, conducted separately at each contrast sensitivity level, revealed no difference between the colour vision testing devices at PRCS ≥1.80 log CS (P ≥ 0.367). However, in all PRCS ≤1.65 log CS, the differences were statistically significant (all P ≤ 0.004), demonstrating a significantly lower percentage of errors in the Ishihara test compared with both the Farnsworth D-15 (P < 0.023) and H-R-R (P < 0.035) tests.

CONCLUSIONS

At high contrast sensitivities, all colour vision tests function almost equally; however, at decreased levels of contrast sensitivity, H-R-R and Farnsworth D-15 are more greatly affected.

摘要

目的

评估对比敏感度降低对三种常用色觉测试的影响,以确定可能与临床实践相关的关键差异。

方法

前瞻性非干预性临床色觉测试研究,使用三种常用设备:石原氏和哈迪-兰德-里特勒(H-R-R)假同色盘测试以及法尔什曼 D-15 排列测试,在逐渐降低对比度敏感度的条件下进行,使用中性密度滤光片条实现。

结果

计算了三种色觉设备中 5%的人群首次经历色觉测试从基线下降 10%时的佩利-罗宾逊对比敏感度(PRCS):法尔什曼 D-15 测试:1.81 对数对比敏感度(CS),H-R-R 测试:1.69 对数 CS,石原氏测试:1.34 对数 CS。在每个对比度敏感度水平上分别进行的单因素重复测量分析显示,在 PRCS ≥1.80 对数 CS 时,三种色觉测试设备之间没有差异(P≥0.367)。然而,在所有 PRCS ≤1.65 对数 CS 时,差异具有统计学意义(均 P≤0.004),表明与法尔什曼 D-15(P<0.023)和 H-R-R(P<0.035)测试相比,石原氏测试的错误率明显更低。

结论

在高对比度敏感度下,所有色觉测试的功能几乎相同;然而,在对比度敏感度降低的情况下,H-R-R 和法尔什曼 D-15 受到的影响更大。

相似文献

1
The effect of reduced contrast sensitivity on colour vision testing.
Eye (Lond). 2019 Jul;33(7):1068-1072. doi: 10.1038/s41433-019-0361-y. Epub 2019 Feb 19.
2
Clinical color vision testing and correlation with visual function.
Am J Ophthalmol. 2015 Sep;160(3):547-552.e1. doi: 10.1016/j.ajo.2015.06.015. Epub 2015 Jun 24.
4
The new Richmond HRR pseudoisochromatic test for colour vision is better than the Ishihara test.
Clin Exp Optom. 2006 Mar;89(2):73-80. doi: 10.1111/j.1444-0938.2006.00015.x.
5
Colour vision testing in young children with reduced visual acuity.
Acta Ophthalmol. 2020 Feb;98(1):e113-e120. doi: 10.1111/aos.14219. Epub 2019 Aug 30.
6
Review of the main colour vision clinical assessment tests.
Arch Soc Esp Oftalmol (Engl Ed). 2019 Jan;94(1):25-32. doi: 10.1016/j.oftal.2018.08.006. Epub 2018 Oct 22.
7
The effect of the ChromaGen contact lens system on visual performance.
Clin Exp Optom. 2020 Jul;103(4):507-512. doi: 10.1111/cxo.13011. Epub 2019 Dec 8.
9
Impact of blur on clinical and occupational colour vision test results.
Ophthalmic Physiol Opt. 2024 Sep;44(6):1058-1071. doi: 10.1111/opo.13350. Epub 2024 Jun 21.
10
Effect of EnChroma glasses on color vision screening using Ishihara and Farnsworth D-15 color vision tests.
J AAPOS. 2020 Jun;24(3):157.e1-157.e5. doi: 10.1016/j.jaapos.2020.03.006. Epub 2020 Jun 1.

引用本文的文献

1
Early predictors of visual and axonal outcomes after acute optic neuritis.
Front Neurol. 2022 Sep 8;13:945034. doi: 10.3389/fneur.2022.945034. eCollection 2022.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验