Suppr超能文献

BD-FACS Presto 在加纳一家地区医院和农村实验室进行 CD4 计数和血红蛋白测量的性能。

Performance of the BD-FACS Presto for CD4 count and hemoglobin measurement in a district hospital and rural laboratory in Ghana.

机构信息

Earth Institute, Columbia University, New York, New York, United States of America.

Millennium Promise Alliance, Accra, Ghana.

出版信息

PLoS One. 2019 Feb 22;14(2):e0212684. doi: 10.1371/journal.pone.0212684. eCollection 2019.

Abstract

INTRODUCTION

In Ghana, initiation of Antiretroviral Therapy (ART) is recommended for all patients with an HIV diagnosis, regardless of CD4+ T-cell count. However, measurement of CD4 count remains an important metric for identifying patients with advanced HIV disease, and assessing a person's overall immune status, which informs the decision to offer opportunistic infection screening and prophylaxis. Access to CD4+ T cell count in rural health facilities remains a major challenge in Ghana and other resource-limited settings. This study aimed to validate the accuracy of the BD FACSPresto near-patient device for measurement of CD4 count and hemoglobin concentration against the FACSCount (CD4) and Sysmex (hemoglobin) diagnostic machines when operated in both a district hospital and rural laboratory, serving a network of health posts in Ashanti Region, Ghana.

METHODOLOGY

In the first phase of the study, patients were recruited from a district hospital, and both venous and capillary blood samples were tested using the FACSCount and Sysmex as reference tests and compared to results of the FACSPresto performed in the clinic laboratory at the district hospital. In the second phase, patients were recruited from both the hospital and from rural health clinics, and samples were tested using the FACSPresto at a rural laboratory. Sensitivity and specificity among samples categorized into different clinically relevant CD4 count ranges were calculated, along with correlation between the Presto and the reference measurements, and mean and relative bias with limits of agreement.

RESULTS

The FACSPresto was successfully operated in both clinical settings. A total of 59 samples in the first phase and 48 samples in the second phase were included. Positive bias was observed when comparing CD4 count measured by BD FACSPresto to FACSCount in the district hospital (bias = 44, LOA -72,160) and in the rural laboratory setting (bias = 74, LOA -96, 244). In addition, capillary blood samples were shown to give higher measures when compared to venous blood samples from the same participant. All results were statistically significant (p<0.05) apart from hemoglobin measurement in venous blood in the rural laboratory. Correlation coefficients were high for CD4 count measures and lower for hemoglobin measures.

CONCLUSION

Overall, the Presto gave higher estimates of CD4 count compared to FACSCount, and hemoglobin measurements were higher than from Sysmex. Samples of capillary blood in turn gave higher results for both measurements compared to venous blood, consistent with previous analyses. These findings should be considered when selecting CD4 count machines for use at the point of care, especially in remote areas where capillary blood sampling may be preferable, but are likely balanced by device's ease of use, portability, and ability to expand access to services. These results are some of the first to demonstrate the accuracy of the FACSPresto in West Africa and show that this device can be successfully operated in a very rural lab setting and may therefore assist to provide CD4 count and hemoglobin concentration measurement to populations in need.

摘要

简介

在加纳,无论 CD4+T 细胞计数如何,所有 HIV 感染者都应开始接受抗逆转录病毒治疗(ART)。然而,CD4 计数的测量仍然是识别晚期 HIV 疾病患者的重要指标,并评估个体的整体免疫状况,这为提供机会性感染筛查和预防提供了依据。在加纳和其他资源有限的地区,农村卫生机构获得 CD4+T 细胞计数仍然是一个主要挑战。本研究旨在验证 BD FACSPresto 即时检测设备在地区医院和农村实验室中测量 CD4 计数和血红蛋白浓度的准确性,该设备用于服务阿散蒂地区的卫生所网络,作为 FACSCount(CD4)和 Sysmex(血红蛋白)诊断仪的替代检测设备。

方法

在研究的第一阶段,从地区医院招募患者,使用 FACSCount 和 Sysmex 作为参考检测方法,同时使用静脉血和毛细血管血样对患者进行检测,并将检测结果与 FACSPresto 在地区医院的诊所实验室中的检测结果进行比较。在第二阶段,从医院和农村诊所招募患者,使用 FACSPresto 在农村实验室进行检测。计算不同临床相关 CD4 计数范围内样本的敏感性和特异性,以及 Presto 与参考测量值之间的相关性,以及均值和相对偏差与一致性界限。

结果

FACSPresto 在两种临床环境中均成功运行。第一阶段共纳入 59 例样本,第二阶段共纳入 48 例样本。与 FACSCount 相比,BD FACSPresto 在地区医院(偏差=44,LOA-72,160)和农村实验室设置(偏差=74,LOA-96,244)中测量 CD4 计数时存在正偏差。此外,与同一参与者的静脉血样本相比,毛细血管血样本的测量值更高。除了农村实验室的静脉血血红蛋白测量外,所有结果均具有统计学意义(p<0.05)。CD4 计数测量的相关系数较高,而血红蛋白测量的相关系数较低。

结论

总体而言,与 FACSCount 相比,Presto 对 CD4 计数的估计值更高,血红蛋白测量值也高于 Sysmex。与静脉血相比,毛细血管血样本的两种测量值都更高,这与之前的分析结果一致。在选择用于即时护理的 CD4 计数设备时,应考虑这些发现,尤其是在偏远地区,那里可能更倾向于进行毛细血管血样采集,但设备的易用性、便携性和扩大服务获取能力可能会对此进行平衡。这些结果是首次在西非证明 FACSPresto 准确性的结果之一,并表明该设备可以在非常偏远的实验室环境中成功运行,因此可能有助于为有需要的人群提供 CD4 计数和血红蛋白浓度测量。