Hwang Jongmin, Kim Jun, Choi Kee Joon, Cho Min Soo, Nam Gi Byoung, Kim You Ho
Division of Cardiology, Department of Internal Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea.
Division of Cardiology, Department of Internal Medicine, Keimyung University Dongsan Medical Center, Daegu, Korea.
Korean Circ J. 2019 May;49(5):437-445. doi: 10.4070/kcj.2018.0323. Epub 2019 Feb 11.
Wrist-worn wearable devices provide heart rate (HR) monitoring function via photoplethysmography technology. Recently, these devices have been used by patients to measure the HR when palpitation occurs, but few validation studies of these instruments have been conducted. We assessed the accuracy of these devices for measuring a HR.
This study enrolled 51 consecutive patients with a history of paroxysmal supraventricular tachyarrhythmia (SVT) or paroxysmal palpitations who were scheduled to undergo an electrophysiological study (EPS). Three devices were assessed: Apple Watch Series 2 (Apple), Samsung Galaxy Gear S3 (Galaxy), and Fitbit Charge 2 (Fitbit). Patients were randomly assigned to wear 2 different devices. The HR at baseline and induced SVT were measured during the EPS. After successful ablation of SVT, HR measurements was also done during atrial and ventricular pacing study.
The mean patient age was 44.4±16.6 years and 27 patients were male (53%). The accuracy (within ±5 beats per minute [bpm] of an electrocardiogram [ECG] measurement) of the baseline HR measurements was 100%, 100%, and 94%, for Apple, Galaxy, and Fitbit, respectively. The HR during induced SVT ranged from 108 bpm to 228 bpm and the accuracy (within ±10 bpm of an ECG) was 100%, 90%, and 87% for the Apple, Galaxy, and Fitbit, respectively. During pacing study, accuracy of these devices was also acceptable but tended to decrease as the HR increased, and showed differences between the devices.
Wrist-worn wearable devices accurately measure baseline and induced SVT HR.
Clinical Research Information Service Identifier: KCT0002282.
腕戴式可穿戴设备通过光电容积脉搏波描记术技术提供心率(HR)监测功能。最近,这些设备已被患者用于在心悸发作时测量心率,但对这些仪器的验证研究很少。我们评估了这些设备测量心率的准确性。
本研究纳入了51例有阵发性室上性快速心律失常(SVT)或阵发性心悸病史且计划接受电生理研究(EPS)的连续患者。评估了三种设备:苹果手表Series 2(Apple)、三星Galaxy Gear S3(Galaxy)和Fitbit Charge 2(Fitbit)。患者被随机分配佩戴2种不同的设备。在EPS期间测量基线心率和诱发SVT时的心率。在成功消融SVT后,还在心房和心室起搏研究期间进行心率测量。
患者的平均年龄为44.4±16.6岁,27例患者为男性(53%)。对于Apple、Galaxy和Fitbit,基线心率测量的准确性(在心电图[ECG]测量的每分钟±5次心跳[bpm]范围内)分别为100%、100%和94%。诱发SVT期间的心率范围为108 bpm至228 bpm,对于Apple、Galaxy和Fitbit,准确性(在ECG的±10 bpm范围内)分别为100%、90%和87%。在起搏研究期间,这些设备的准确性也可以接受,但随着心率增加有下降趋势,并且设备之间存在差异。
腕戴式可穿戴设备能准确测量基线心率和诱发SVT时的心率。
临床研究信息服务标识符:KCT0002282。