Jensen J M, Brix F, Kohr P
Strahlenther Onkol. 1986 Apr;162(4):250-3.
In magna-field irradiation using high energy bremsstrahlung (photons), i.e. total body irradiation, it is desirable to achieve a homogeneous dose distribution or to realize a planned modification in dose deposition within the target volume. In the Department of Radiology, University of Kiel, for this purpose individually constructed compensators are in routine use for about three years. Some physical and dosimetric aspects of this irradiation technique are described briefly and the experimental test procedure for TBI-compensators is highlighted by several details, for example transmission measurements, radiographies of the compensators and determination of actual dose rates in the midplane of the patient.
在使用高能轫致辐射(光子)的大野照射中,即全身照射,期望实现均匀的剂量分布或在靶区内实现计划的剂量沉积修正。在基尔大学放射科,为此目的,定制的补偿器已常规使用约三年。简要描述了这种照射技术的一些物理和剂量学方面,并通过几个细节突出了全身照射补偿器的实验测试程序,例如透射测量、补偿器的射线照相以及患者中平面实际剂量率的测定。