对比增强和图像质量对二维和三维超声心动图测定左心室容积的影响:与磁共振成像的比较

Contrast Enhancement and Image Quality Influence Two- and Three-dimensional Echocardiographic Determination of Left Ventricular Volumes: Comparison With Magnetic Resonance Imaging.

作者信息

Jenner Jonas, Sörensson Peder, Pernow John, Caidahl Kenneth, Eriksson Maria J

机构信息

Department of Molecular Medicine and Surgery, Karolinska Institutet, Stockholm, Sweden.

Department of Clinical Physiology, Karolinska University Hospital, Stockholm, Sweden.

出版信息

Clin Med Insights Cardiol. 2019 Mar 5;13:1179546819831980. doi: 10.1177/1179546819831980. eCollection 2019.

Abstract

PURPOSE

To evaluate the effect of image quality and contrast enhancement (CE) on left ventricular (LV) volume determination by two- (2D) and three-dimensional (3D) echocardiography (2DE/3DE).

METHODS

We studied 32 post-myocardial infarction (MI) patients without (2DE/3DE) and with CE (CE2DE/CE3DE), in comparison with cardiac magnetic resonance imaging (CMR).

RESULTS

Two-dimensional echocardiography showed the largest negative bias versus CMR for diastolic and systolic volumes (-59, -28 mL, respectively) with lower biases for CE2DE (-37, -22 mL), 3DE (-31, -17 mL), and CE3DE (-17, -11 mL). Bias for ejection fraction (EF) ranged from -2.1% for 2DE to +1.4% for CE3DE. Agreement (intraclass correlation coefficient, ICC) for EF between CMR and 3DE (0.86 without and 0.85 with contrast) was better than for 2DE (0.73 without and 0.69 with contrast). The inter-/intra-observer coefficients of variation for EF varied from 16%/10% (2DE) to 6.9%/6.6% (CE2DE), and 8.3%/4.8% (3DE) to 6.7%/6.8% (CE3DE), respectively. The agreement (ICC) with CMR for EF measured by 2DE/3DE changed from 0.64/0.84 with poor image quality to 0.81/0.87 with moderate to good image quality.

CONCLUSIONS

Three-dimensional echocardiography was more accurate than 2DE for estimating LV volumes, with less inter-/intra-observer variability in EF values. Contrast enhancement improved accuracy for both 2DE and 3DE and improved the inter-observer variability of EF estimates for 2DE and 3DE. Image quality had more impact on the agreement of EF values with CMR for 2DE than for 3DE. Our results emphasize the importance of using the same technique for longitudinal studies of LV EF and specially LV volumes.

摘要

目的

评估图像质量和对比增强(CE)对二维(2D)和三维(3D)超声心动图(2DE/3DE)测定左心室(LV)容积的影响。

方法

我们研究了32例心肌梗死(MI)后患者,分别进行了无CE(2DE/3DE)和有CE(CE2DE/CE3DE)检查,并与心脏磁共振成像(CMR)进行比较。

结果

二维超声心动图显示,与CMR相比,舒张期和收缩期容积的负偏差最大(分别为-59、-28 mL),CE2DE(-37、-22 mL)、3DE(-31、-17 mL)和CE3DE(-17、-11 mL)的偏差较小。射血分数(EF)的偏差范围从2DE的-2.1%到CE3DE的+1.4%。CMR与3DE之间EF的一致性(组内相关系数,ICC)(无对比时为0.86,有对比时为0.85)优于2DE(无对比时为0.73,有对比时为0.69)。EF的观察者间/观察者内变异系数分别从16%/10%(2DE)到6.9%/6.6%(CE2DE),以及8.3%/4.8%(3DE)到6.7%/6.8%(CE3DE)。2DE/3DE测量的EF与CMR的一致性(ICC)从图像质量差时的0.64/0.84变为图像质量中等至良好时的0.81/0.87。

结论

三维超声心动图在估计LV容积方面比2DE更准确,EF值的观察者间/观察者内变异性更小。对比增强提高了2DE和3DE的准确性,并改善了2DE和3DE的EF估计值的观察者间变异性。图像质量对2DE的EF值与CMR一致性的影响比对3DE的影响更大。我们的结果强调了在LV EF特别是LV容积的纵向研究中使用相同技术的重要性。