[123I]FP-CIT单光子发射计算机断层扫描中特异性结合率的标准化:纹状体模型研究

Standardization of the specific binding ratio in [123I]FP-CIT SPECT: study by striatum phantom.

作者信息

Kita Akinobu, Onoguchi Masahisa, Shibutani Takayuki, Horita Hirokazu, Oku Yasuhiro, Kashiwaya Souichirou, Isaka Masato, Saitou Masaki

机构信息

Radiological Center, University of Fukui Hospital, Fukui.

Department of Quantum Medical Technology, Graduate School of Medical Sciences.

出版信息

Nucl Med Commun. 2019 May;40(5):484-490. doi: 10.1097/MNM.0000000000001010.

Abstract

OBJECTIVE

The quantitative values of the specific binding ratio (SBR) in [I]FP-CIT have been reported to change because of differences in apparatus, collection conditions, and image reconstruction. The aim of this study was to clarify the distribution of calculated SBR values by performing [I]FP-CIT single-photon emission computed tomography in a multicenter collaborative study using a phantom. A simple correction method was also devised that enables direct comparison of the SBR value calculated at one facility with those calculated at other facilities.

MATERIALS AND METHODS

Data were acquired at 14 facilities using the phantom adjusted to right striatum : left striatum : cerebral parenchyma (back ground)=8 : 4 : 1, and the SBR values were calculated.We devised a method to correct the SBR using the results of experiments with a known ratio phantom.

RESULTS

The SBR values considerably differed between facilities. The average SBR with a theoretical value of 7 and with a theoretical value of 3 in all facilities was 6.48±0.89 and 2.58±0.51, respectively. The range of SBRs with a theoretical value of 7 and a theoretical value of 3 was 3.18 and 1.59, respectively. We devised a simple method for calibrating the SBR value at the clinical examination of each facility to a directly comparable value.

CONCLUSION

Direct comparison of the SBR with those of other facilities and sharing other facilities normal values is clinically difficult. We devised countermeasures that do not affect the diagnosis and developed a simple tool to calculate the standardized SBR.

摘要

目的

据报道,由于仪器、采集条件和图像重建的差异,[I]FP - CIT中特异性结合率(SBR)的定量值会发生变化。本研究的目的是通过在多中心协作研究中使用体模进行[I]FP - CIT单光子发射计算机断层扫描,阐明计算得到的SBR值的分布情况。还设计了一种简单的校正方法,能够直接比较在一个机构计算得到的SBR值与在其他机构计算得到的SBR值。

材料与方法

在14个机构使用调整为右纹状体∶左纹状体∶脑实质(背景) = 8∶4∶1的体模采集数据,并计算SBR值。我们设计了一种利用已知比例体模的实验结果校正SBR的方法。

结果

各机构之间的SBR值差异很大。在所有机构中,理论值为7和理论值为3时的平均SBR分别为6.48±0.89和2.58±0.51。理论值为7和理论值为3时SBR的范围分别为3.18和1.59。我们设计了一种简单的方法,可将每个机构临床检查时的SBR值校准为可直接比较的值。

结论

临床上很难直接将一个机构的SBR与其他机构的SBR进行比较并共享其他机构的正常值。我们制定了不影响诊断的应对措施,并开发了一种简单的工具来计算标准化SBR。