Kayano Shingo, Takano Hirokazu, Takane Yumi, Satomura Ayaka, Ono Katsunori, Shimura Hirotaka, Onodera Shu, Sato Kazuhiro
Department of Radiological Technology, Tohoku University Hospital.
Health Sciences, Tohoku University Graduate School of Medicine.
Nihon Hoshasen Gijutsu Gakkai Zasshi. 2019;75(3):247-253. doi: 10.6009/jjrt.2019_JSRT_75.3.247.
Dual-energy computed tomography (DE-CT) is the promising technology, such as enabling material decomposition, generation of the virtual monochromatic image, and measurement of effective atomic numbers. There are reports that utilization of the virtual non-contrast (VNC) image, the iodine map image, and the virtual monochromatic image can contribute to the improvement of lesion detection and its characterization, compared with conventional contrast CT by single-energy computed tomography (SE-CT). In addition, acquisition of the VNC images makes it possible to skip scanning of true non-contrast CT, which is also expected to reduce exposure. However, a reliable evaluation of the accuracy of the VNC image has not been established, and only a few reports have verified their accuracy. In this study, we evaluated the relationship between the quantitativeness of iodine and the CT value of VNC image. As a result of our study, when the iodine volume was overestimated, the CT value of the VNC image was lower than the reference value, and when the iodine volume was underestimated, the CT value was upper than the reference value. Moreover, we clarified that the CT value of the VNC image greatly diverges as the iodine volume increases.
双能计算机断层扫描(DE-CT)是一项很有前景的技术,例如能够实现物质分解、生成虚拟单色图像以及测量有效原子序数。有报告称,与传统的单能计算机断层扫描(SE-CT)对比CT相比,利用虚拟平扫(VNC)图像、碘图图像和虚拟单色图像有助于提高病变检测及其特征描述。此外,获取VNC图像使得跳过真正的平扫CT扫描成为可能,这也有望减少辐射剂量。然而,尚未建立对VNC图像准确性的可靠评估,仅有少数报告验证了其准确性。在本研究中,我们评估了碘的定量与VNC图像CT值之间的关系。我们的研究结果表明,当碘含量被高估时,VNC图像的CT值低于参考值,而当碘含量被低估时,CT值高于参考值。此外,我们还阐明,随着碘含量的增加,VNC图像的CT值差异很大。