J Cyst Fibros. 2019 Nov;18(6):e56-e57. doi: 10.1016/j.jcf.2019.03.004. Epub 2019 Mar 21.
Fifty-two meshes of e-flow rapid® were characterized for tobramycin delivery with a laser diffractometer after 6 months of home use by cystic fibrosis patients treated with various nebulized drugs. Three meshes were out of order and 30 considered to be defective for tobramycin delivery. The use of the specific mesh cleaning shower system permitted 14 defective meshes to be in the expected range of nebulized volume.
52 个 e-flow rapid® 滤网经激光衍射仪检测,发现 30 个有缺陷,无法输送妥布霉素,这 30 个滤网是在 6 个月的家庭使用期间,由接受不同雾化药物治疗的囊性纤维化患者使用的。另有 3 个滤网出现故障,3 个被认为无法输送妥布霉素。使用专用的滤网清洗淋浴系统后,14 个有缺陷的滤网的雾化体积恢复到预期范围。