Microsc Res Tech. 2019 Jul;82(7):1157-1164. doi: 10.1002/jemt.23264. Epub 2019 Mar 27.
The aim was to examine the effect of retreatment process on the surface roughness and nickel titanium (NiTi) composition of ProTaper Universal Retreatment (PTUR; consists of 3 files; D1, D2, D3) and WaveOne Gold (WOG) (primary) instruments. Twenty extracted mandibular molar teeth with severe curved (30-40°) mesial roots were selected and divided into two groups (n = 10) based on the instrument used for the removal of the root canal filling. Before and after using the instruments in two canals, they were subjected to atomic force microscopy (AFM) and energy dispersive X-ray spectrophotometry (EDX) analysis. The EDX analysis data and roughness average (Ra) and root mean square (RMS) values were analyzed statistically using a one-way analysis of variance and post hoc Tukey's test at the 5% significant level. There was no significant difference between the new and used D1 and D2 PTUR and WOG instruments in terms of the Ni composition (p > .05). The Ti contents of the used D2 and D3 PTUR instruments were lower those of the new instruments (p < .05). In both new and used instruments, PTUR and WOG have no difference in terms of Ra and RMS values. (p > .05). The Ra and RMS values of the PTUR and WOG systems significantly increased after removal of the root canal filling (p < .05). The use of PTUR and WOG instruments for removal of root canal filling in severely curved root canals affected the surface topography of the files. The NiTi composition of the WOG instruments was unaffected by the retreatment process.
目的在于研究再治疗过程对 ProTaper Universal Retreatment(PTUR;由 3 个锉组成,即 D1、D2、D3)和 WaveOne Gold(WOG)(主要锉)器械的表面粗糙度和镍钛(NiTi)成分的影响。选择 20 颗具有严重弯曲(30-40°)近中根的下颌磨牙,并根据根管充填去除所使用的器械将其分为两组(n=10)。在两根管中使用器械前后,对其进行原子力显微镜(AFM)和能谱分析(EDX)。采用单因素方差分析和事后 Tukey 检验,对 EDX 分析数据和粗糙度平均值(Ra)和均方根(RMS)值进行统计学分析,检验水准为 5%。新、用过的 D1 和 D2 PTUR 与 WOG 器械的 Ni 成分无显著差异(p>.05)。用过的 D2 和 D3 PTUR 器械的 Ti 含量低于新器械(p<.05)。新、用过的器械中,PTUR 和 WOG 的 Ra 和 RMS 值无差异(p>.05)。去除根管充填物后,PTUR 和 WOG 系统的 Ra 和 RMS 值显著增加(p<.05)。在严重弯曲根管中使用 PTUR 和 WOG 器械去除根管充填物会影响器械的表面形貌。WOG 器械的 NiTi 成分不受再治疗过程的影响。