Recently developed Nickel-Titanium (NiTi) instruments with practical changes have resulted in safer instrumentation. In addition, topographical features on the file surface are a contributing factor to clinical durability. Therefore, this study aimed to investigate both the cyclic fatigue resistance and the roughness change of MTwo and Rotate instruments (VDW, Munich, Germany). Each instrument (n = 6/each group) was scanned with an atomic force microscopy prior to and after instrumentation. In addition, cyclic fatigue testing was conducted for each instrument (n = 11/each group) with stainless-steel blocks, including 45°-60°-90° degrees of curvature milled to the instruments' size. The roughness parameters increased for both systems after instrumentation (p<0.05). Both systems presented an increased roughness following instrumentation (p<0.05). The cyclic fatigue resistance was lowest at 90° for both systems (p<0.05), whereas the Rotate files presented a higher resistance than that of the Mtwo files (p<0.05). Compared to the Mtwo files, Rotate files presented better resistance, while the resistance decreased as the curvature increased.
最近开发的具有实际改进的镍钛(NiTi)器械导致了更安全的器械。此外,锉表面的形貌特征是临床耐用性的一个影响因素。因此,本研究旨在研究 MTwo 和 Rotate 器械(VDW,慕尼黑,德国)的循环疲劳阻力和粗糙度变化。在器械处理前后,用原子力显微镜对每个器械(每组 6 个)进行扫描。此外,对每个器械(每组 11 个)进行循环疲劳测试,使用不锈钢块,将器械尺寸铣成 45°-60°-90°的曲率。器械处理后,两个系统的粗糙度参数均增加(p<0.05)。两个系统在器械处理后粗糙度均增加(p<0.05)。两种系统的 90°的循环疲劳阻力最低(p<0.05),而 Rotate 锉的阻力高于 Mtwo 锉(p<0.05)。与 Mtwo 锉相比,Rotate 锉具有更好的阻力,而阻力随着曲率的增加而降低。