Swansea University - Diabetes Research Group, Swansea, UK.
Diabet Med. 2019 Jul;36(7):862-867. doi: 10.1111/dme.13961. Epub 2019 Apr 26.
To evaluate the performance of the current, pre-production version of a novel home oral glucose tolerance test (Home OGTT) device when administered by trained research nurses, compared with a reference laboratory glucose analyser and a second laboratory analyser, incorporating a sample processing delay to simulate normal practice.
One hundred women (aged 19-48 years), with and without known glucose intolerance were recruited. Following an overnight fast, participants attended for a 75-g OGTT. A fasting capillary sample was applied to the Home OGTT device with a corresponding venous sample collected and measured immediately on the reference YSI 2300 stat plus analyser, and following a 1-h delay on the Randox Daytona Plus analyser. The sampling process was repeated 2 h after the oral glucose load.
Some 97% of tested devices gave complete data for analysis. Good agreement was observed between the reference glucose analyser and the Home OGTT device, with the Home OGTT device displaying a small negative bias (-0.18 mmol/l, -1.75 to 1.39 mmol/mol; -1.0%, -26.4% to 24.5%; absolute and relative mean, 95% limits of agreement). When classified as normal glucose tolerant or glucose intolerant, the Home OGTT device showed 100% and 90% sensitivity, and 99% and 99% specificity using fasting plasma glucose and 2-h glucose respectively. Similar sensitivity (100% and 100%) and specificity (96% and 99%) for fasting plasma glucose and 2-h glucose were observed using the secondary analyser.
The novel Home OGTT device was reliable and easy to use and showed excellent agreement with two separate laboratory analysers. The Home OGTT offers potential as an effective alternative for clinic-based OGTT testing.
评估一种新型家庭口服葡萄糖耐量试验(Home OGTT)设备在经过培训的研究护士使用时的性能,与参考实验室葡萄糖分析仪和第二实验室分析仪进行比较,同时模拟正常实践中的样本处理延迟。
共招募了 100 名年龄在 19-48 岁之间的女性,包括已知糖耐量受损和非已知糖耐量受损的女性。禁食过夜后,参与者进行 75g OGTT。将空腹毛细血管样本应用于 Home OGTT 设备,同时采集相应的静脉样本,并立即在参考 YSI 2300 stat plus 分析仪上进行测量,在 Randox Daytona Plus 分析仪上延迟 1 小时后再次进行测量。在口服葡萄糖负荷后 2 小时重复采样过程。
约 97%的测试设备提供了完整的数据进行分析。参考葡萄糖分析仪与 Home OGTT 设备之间观察到良好的一致性,Home OGTT 设备显示出小的负偏差(-0.18mmol/l,-1.75 至 1.39mmol/mol;-1.0%,-26.4%至 24.5%;绝对和相对均值,95%一致性界限)。当根据空腹血糖和 2 小时血糖将其分类为正常糖耐量或糖耐量受损时,Home OGTT 设备分别显示出 100%和 90%的敏感性,以及 99%和 99%的特异性。使用二级分析仪时,空腹血糖和 2 小时血糖的敏感性(100%和 100%)和特异性(96%和 99%)也相似。
新型 Home OGTT 设备可靠且易于使用,与两个独立的实验室分析仪具有极好的一致性。Home OGTT 作为一种有效的诊所 OGTT 检测替代方法具有潜力。