J Magn Reson Imaging. 2019 Nov;50(5):1614-1619. doi: 10.1002/jmri.26751. Epub 2019 Apr 16.
Evaluation of prostate MRI relies on diffusion-weighted imaging (DWI), commonly distorted by susceptibility artifacts, thereby creating a need for approaches to correct such distortion.
To compare geometric distortion on prostate MRI between standard DWI and a geometric distortion correction method for DWI described as static distortion correction DWI (SDC DWI).
Retrospective case study.
Thirty patients (ages 31-81 years) undergoing prostate MRI.
Geometric distortions from echo planar imaging were corrected with the SDC DWI protocol, which first acquires a B -field map to estimate geometric distortions.
Contours of the prostate were placed on axial T -weighted imaging (T WI) as an anatomic standard. Pixel shifts and apparent diffusion coefficient (ADC) values were compared between prostate contours applied to the SDC DWI and standard DWI sequences. Detailed characterization of the impact of SDC DWI was performed in three representative patients.
One-way analysis of variance (ANOVA) test, Spearman correlation test, and Bland-Altman plots were calculated.
There was significantly greater overlap of the SDC DWI prostate region of interest (ROI) with T WI than standard DWI with T WI (10.56 cm ± 3.14, P < 0.05). R of ADC values from standard DWI vs. SDC DWI in the 30 patients ranged from 0.02-0.94 (mean 0.60). A patient without susceptibility artifact demonstrated minimal pixel shift ranging from 0.6-1.3 mm and high correlation of ADC values (R = 0.89) between SDC DWI and standard DWI. A patient with rectal gas showed greater pixel shift (range: -2.5 to -0.5 mm) and less ADC value correlation (R = 0.69). A patient with a pelvic phlebolith adjacent to the prostate showed an even greater pixel shift (range: 10-16 mm) and decreased ADC correlation (R = 0.21).
SDC DWI appears to correct for susceptibility-related pixel shifts in the prostate compared with standard DWI, which may have value for assessing prostate lesions obscured by geometric warping. Level of Evidence 4 Technical Efficacy Stage 1 J. Magn. Reson. Imaging 2019;50:1614-1619.
前列腺 MRI 的评估依赖于扩散加权成像(DWI),但 DWI 通常会受到磁化率伪影的扭曲,因此需要采用校正这些扭曲的方法。
比较标准 DWI 和一种称为静态失真校正 DWI(SDC DWI)的 DWI 几何失真校正方法在前列腺 MRI 中的几何失真。
回顾性病例研究。
30 名(年龄 31-81 岁)行前列腺 MRI 的患者。
采用 SDC DWI 方案校正 EPI 中的几何失真,该方案首先获取 B 场图以估计几何失真。
在轴位 T 加权成像(TWI)上绘制前列腺轮廓作为解剖标准。比较应用于 SDC DWI 和标准 DWI 序列的前列腺轮廓的像素移位和表观扩散系数(ADC)值。在 3 名代表性患者中详细描述 SDC DWI 的影响。
采用单因素方差分析(ANOVA)检验、Spearman 相关检验和 Bland-Altman 图进行计算。
与 TWI 相比,SDC DWI 的前列腺感兴趣区(ROI)重叠度明显更高(10.56cm±3.14,P<0.05)。30 名患者中标准 DWI 与 SDC DWI 的 ADC 值 R 范围为 0.02-0.94(均值 0.60)。无磁化率伪影的患者的像素移位范围为 0.6-1.3mm,ADC 值相关性较高(R=0.89)。有直肠气体的患者的像素移位较大(范围:-2.5 至-0.5mm),ADC 值相关性较低(R=0.69)。有靠近前列腺的盆腔静脉石的患者的像素移位更大(范围:10-16mm),ADC 值相关性降低(R=0.21)。
与标准 DWI 相比,SDC DWI 似乎可以校正前列腺的磁化率相关像素移位,这对于评估受几何变形影响的前列腺病变可能具有价值。
磁共振成像杂志 2019 年 50 卷 1614-1619 页
证据水平 4 技术功效阶段 1