因金属过敏导致的内侧单髁膝关节置换失败

Failure of a Medial Unicompartmental Knee Replacement due to Metal Allergy.

作者信息

Apostolopoulos Alexandros P, Balfousias Theodore, Khan Shamsul, Srikanth Bala, Tselentakis George

机构信息

Trauma and Orthopaedic Department, Prince Charles Hospital, London, UK.

Trauma and Orthopaedic Department, East Surrey Hospital, Surrey and Sussex Healthcare, NHS Trust, Surrey United Kingdom RH15RH.

出版信息

J Long Term Eff Med Implants. 2018;28(4):319-325. doi: 10.1615/JLongTermEffMedImplants.2019030060.

Abstract

Metal hypersensitivity is an uncommon complication of knee arthroplasty that can lead to significant functional impairment and aseptic prosthesis failure. We describe the case of a 65-yr-old female patient who presented with persistent pain, swelling, and instability 1 yr after medial unicompartmental knee arthroplasty (UKA). The patient had no history of trauma or fall. The initial clinical examination showed good range of movement in the knee joint and no signs of infection or inflammation. Initial radiographs did not reveal aseptic loosening of the implant. Inflammatory markers were normal and the patient was sent for a bone scintigraphy that revealed increased uptake of the femoral implant. Pain deteriorated during the proceeding weeks, and repeated radiographs revealed loosening and dislocation of the femoral implant. We performed a nickel lymphocyte proliferation skin allergy test that resulted in a positive reaction. Revision total knee arthroplasty was performed using an oxynium (zirconium metal alloy) implant with intraoperative tissue pathology and postoperative leukocyte transformation testing that confirmed metal hypersensitivity as the cause for aseptic implant failure. This case report demonstrates that clinical and laboratory signs suggesting metal hypersensitivity in knee arthroplasty can suggest early loosening and failure of the implant.

摘要

金属超敏反应是膝关节置换术一种罕见的并发症,可导致严重的功能障碍和无菌性假体失败。我们描述了一名65岁女性患者的病例,该患者在内侧单髁膝关节置换术(UKA)后1年出现持续疼痛、肿胀和不稳定。患者无外伤或跌倒史。最初的临床检查显示膝关节活动范围良好,无感染或炎症迹象。最初的X线片未显示植入物无菌性松动。炎症标志物正常,患者接受了骨闪烁显像检查,结果显示股骨植入物摄取增加。在接下来的几周内疼痛加剧,重复的X线片显示股骨植入物松动和脱位。我们进行了镍淋巴细胞增殖皮肤过敏试验,结果呈阳性反应。使用钇稳定氧化锆(锆金属合金)植入物进行了翻修全膝关节置换术,术中组织病理学检查和术后白细胞转化试验证实金属超敏反应是无菌性植入物失败的原因。本病例报告表明,膝关节置换术中提示金属超敏反应的临床和实验室体征可提示植入物早期松动和失败。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验