Abstract

Lymphocyte transformation testing (LTT) is often used in the workup for possible metal allergy after total knee arthroplasty (TKA) but the correlation of this test with other diagnostic metal-allergy findings in patients undergoing revision TKA for suspected metal allergy has not been established. A single-center, single-surgeon cohort of 19 TKAs in which both components were revised for presumed implant-related metal allergy based on history, physical, and LTT testing, to nonnickel-containing implants were retrospectively identified. Histopathologic samples obtained intraoperatively were semiquantitatively analyzed using both the Hospital for Special Surgery (HSS) synovial pathology score and the Campbell aseptic lymphocyte-dominant vasculitis-associated lesion (ALVAL) score. As histopathology control group, we included in the study an additional cohort of 17 patients who received aseptic revision TKA and had no history of reported or tested metal sensitivity. All preoperative LTT results were highly reactive to nickel. However, this did not correlate with local periarticular tissue response in 18 of 19 cases which demonstrated a low HSS synovial score (mean: 3.8 ± 2.8, of a maximum score of 28) and the low Campbell ALVAL scores (mean: 2.5/10 ± 1.3, of a maximum score of 10). There were not any significant differences between the study group (suspected implant-related metal allergy) and the control group (nonsuspected implant-related metal allergy) in regard to (1) the Campbell score and (2) the HSS synovial inflammatory score. Knee Society Clinical Rating System (KSCRS) function score improved significantly after revision (mean postoperative increase: 34.0 ± 17. 2;  < 0.001), as well as mean visual analog scale (VAS) pain (mean postoperative decrease: 33.3 ± 26.4;  < 0.01) score. The short-term survival rate (at mean follow-up of 26.1 months) of this patient cohort was 100%. In this cohort of revised TKA patients with suspected nickel allergy based on clinical presentation and LTT positive results, intraoperative histopathology was essentially normal. However, all patients with suspected nickel allergy showed a significant clinical and functional improvement with excellent short-term survival rates. The clinical significance of a positive LTT needs further study.

摘要

淋巴细胞转化试验(LTT)常用于全膝关节置换术后(TKA)可能的金属过敏检测,但这种检测与接受可疑金属过敏的 TKA 翻修术患者的其他诊断性金属过敏发现之间的相关性尚未确定。本研究回顾性分析了 19 例 TKA,这些 TKA 因疑似植入物相关金属过敏,基于病史、体格检查和 LTT 检查,对所有组件均进行了翻修,使用非含镍植入物。术中获得的组织病理学样本使用特殊外科医院(HSS)滑膜炎病理评分和 Campbell 无菌淋巴细胞占主导的血管炎相关病变(ALVAL)评分进行半定量分析。作为组织病理学对照组,我们纳入了另外 17 例接受无菌翻修 TKA 且无报告或检测到金属敏感性病史的患者。所有术前 LTT 结果均对镍高度反应,但这与 18 例中的 19 例局部关节周围组织反应没有相关性,这 18 例患者的 HSS 滑膜炎评分较低(均值:3.8±2.8,最高评分为 28),Campbell ALVAL 评分较低(均值:2.5/10±1.3,最高评分为 10)。在研究组(疑似植入物相关金属过敏)和对照组(非疑似植入物相关金属过敏)之间,(1)Campbell 评分和(2)HSS 滑膜炎炎症评分均无显著差异。翻修后膝关节学会临床评分系统(KSCRS)功能评分显著改善(平均术后增加:34.0±17.2;<0.001),平均视觉模拟评分(VAS)疼痛评分也显著降低(平均术后降低:33.3±26.4;<0.01)。该患者队列的短期生存率(平均随访 26.1 个月)为 100%。在这个基于临床表现和 LTT 阳性结果的疑似镍过敏的 TKA 翻修患者队列中,术中组织病理学基本正常。然而,所有疑似镍过敏的患者均表现出显著的临床和功能改善,且短期生存率非常高。LTT 阳性的临床意义需要进一步研究。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验