The impact of glistenings on the optical quality of a hydrophobic acrylic intraocular lens.

机构信息

David J. Apple International Laboratory for Ocular Pathology, Department of Ophthalmology, University of Heidelberg, Germany.

David J. Apple International Laboratory for Ocular Pathology, Department of Ophthalmology, University of Heidelberg, Germany.

出版信息

J Cataract Refract Surg. 2019 Jul;45(7):1020-1025. doi: 10.1016/j.jcrs.2019.01.025. Epub 2019 Apr 16.

Abstract

PURPOSE

To study the effect of glistenings on the optical quality of a hydrophobic acrylic intraocular lens (IOL).

SETTING

David J. Apple Laboratory, Heidelberg, Germany.

DESIGN

An in vitro laboratory study.

METHODS

An accelerated aging protocol was used to induce glistenings (microvacuoles) in 38 hydrophobic acrylic IOLs. The IOLs were warmed to 45°C and then cooled to 37°C. Image analysis of light-microcopy photographs determined the number and size of microvacuoles (MV). A classification was applied based on the glistening number per mm: grade 0 (none), grade 1 (1-100), grade 2 (101-200), grade 3 (201-500) and grade 4 (more than 500). An optical bench was used to measure each IOL's modulation transfer function (MTF) and Strehl ratio to evaluate the impact of glistenings on image quality.

RESULTS

Glistenings were observed in all IOLs. The mean glistening numbers ± SD (MV/mm) in grades 1 through 4 were 74 ± 12.7, 142 ± 22.2, 297 ± 76.2, and 1509 ± 311.9, respectively. The mean glistening sizes in grades 1 through 4 were 13.28 ± 3.85 μm, 15.88 ± 2.08 μm, 16.85 ± 3.23 μm, and 15.27 ± 2.25 μm, respectively. Statistical analysis showed that grades 1 through 3 did not change the optical quality; however, in grade 4, the MTF and the Strehl ratio were significantly affected.

CONCLUSION

A limited number of glistenings (<500 MV/mm) had no effect on the image quality. Although statistically significant deteriorations in the MTF and the Strehl ratio were observed in grade 4, the effects found were small and are unlikely to affect the visual quality.

摘要

目的

研究蛋白多糖对疏水性丙烯酸人工晶状体(IOL)光学质量的影响。

地点

德国海德堡的 David J. Apple 实验室。

设计

体外实验室研究。

方法

采用加速老化方案在 38 个疏水性丙烯酸 IOL 中诱导蛋白多糖(微囊泡)。将 IOL 加热至 45°C,然后冷却至 37°C。通过光镜照片的图像分析确定微囊泡(MV)的数量和大小。根据每毫米的蛋白多糖数量进行分级:0 级(无)、1 级(1-100)、2 级(101-200)、3 级(201-500)和 4 级(>500)。使用光学台测量每个 IOL 的调制传递函数(MTF)和斯特列尔比,以评估蛋白多糖对图像质量的影响。

结果

所有 IOL 中均观察到蛋白多糖。1 级至 4 级的平均蛋白多糖数量(MV/mm)分别为 74±12.7、142±22.2、297±76.2 和 1509±311.9。1 级至 4 级的平均蛋白多糖大小分别为 13.28±3.85μm、15.88±2.08μm、16.85±3.23μm 和 15.27±2.25μm。统计学分析表明,1 级至 3 级不会改变光学质量;然而,在 4 级,MTF 和斯特列尔比受到显著影响。

结论

数量有限的蛋白多糖(<500 MV/mm)不会影响图像质量。尽管在 4 级观察到 MTF 和斯特列尔比的统计学显著恶化,但发现的影响很小,不太可能影响视觉质量。