Echocardiography. 2019 May;36(5):944-947. doi: 10.1111/echo.14335. Epub 2019 Apr 22.

Abstract

OBJECTIVES

To report the usefulness of harmonic imaging in echocardiography to visualize and direct the implantation of the Medtronic micro vascular plug (MVP).

BACKGROUND

Off label use of the MVP was reported for transcatheter occlusion of patent ductus arteriosus (PDA) in premature infants. The device is poorly visible on fluoroscopy and echocardiography.

METHODS

In 9 consecutive premature infants, the MVP was used for transcatheter closure of the PDA. In each, the ability of conventional echocardiographic imaging was compared to harmonic imaging, and the device was deployed in the PDA using echocardiography.

RESULTS

In each subject, harmonic imaging proved superior to conventional echocardiography to visualize the MVP in premature infants using 12 and 8 MHz probes. Once the delivery, catheter was across the PDA into the descending aorta, and the MVP advanced to the catheter tip, positioning, and deployment of the device was possible without fluoroscopy. All devices were deployed appropriately with immediate occlusion and no obstruction to the left pulmonary artery or aorta.

CONCLUSIONS

The MVP can be accurately imaged using harmonic imaging, even in the near field in premature infants. Precise implantation of the MVP in the PDA of premature infants is possible with echocardiographic imaging of the device and vascular structures.

摘要

目的

报告谐波成像在超声心动图中对可视化和指导美敦力微血管封堵器(MVP)植入的作用。

背景

据报道,MVP在早产儿动脉导管未闭(PDA)经导管封堵中存在未按标签说明使用的情况。该装置在荧光透视和超声心动图上显示不佳。

方法

对9例连续的早产儿使用MVP进行PDA经导管封堵。在每例患儿中,将传统超声心动图成像与谐波成像的能力进行比较,并使用超声心动图将装置部署于PDA中。

结果

在每个研究对象中,使用12兆赫和8兆赫探头时,谐波成像在可视化早产儿的MVP方面优于传统超声心动图。一旦输送导管穿过PDA进入降主动脉,且MVP推进到导管尖端,无需荧光透视即可进行装置的定位和部署。所有装置均正确部署,立即实现封堵,且未对左肺动脉或主动脉造成梗阻。

结论

即使在早产儿的近场中,使用谐波成像也可准确显示MVP。通过对装置和血管结构进行超声心动图成像,可在早产儿的PDA中精确植入MVP。