A uniaxial load cell device provides an alternative, easy and inexpensive way to quantify muscle strength in different settings outside the clinic and research labs. So, the purpose of the study was to examine the test-retest reliability and the construct validity of lower extremity strength performance using an uniaxial load cell device.
A total of 131 subjects (85% female, mean age 80 ± 8 years) were included for the validity aim, and a sample of 38 subjects were enrolled in the reliability testing (89% female, mean age 76 ± 7 years). For the strength measurements were assessed with a portable load cell for three consecutive trials. Test-retest reliability was assessed over two testing visits occurring one week apart. Spearman's rank correlation coefficient was used to test convergent validity with other mobility-related measurements construct validity at baseline.
Strength measurements showed good to excellent reliability in most of the measured parameters with intraclass correlation coefficients range from 0.89 to 0.99 and were correlated with mobility measurements with Spearman rho range from 0.21 to 0.38.
The portable uni-axial load cell to measure lower extremity strength provides reliable measurements in community settings.
单轴测力传感器装置提供了一种在诊所和研究实验室之外的不同环境中量化肌肉力量的替代方法,既简便又经济。因此,本研究的目的是使用单轴测力传感器装置检验下肢力量表现的重测信度和结构效度。
共有131名受试者(85%为女性,平均年龄80±8岁)纳入效度研究,38名受试者样本纳入信度测试(89%为女性,平均年龄76±7岁)。使用便携式测力传感器对力量测量进行连续三次试验评估。重测信度在相隔一周的两次测试访问中进行评估。使用Spearman等级相关系数在基线时测试与其他移动性相关测量的收敛效度和结构效度。
力量测量在大多数测量参数中显示出良好到极好的信度,组内相关系数范围为0.89至0.99,并且与移动性测量相关,Spearman相关系数范围为0.21至0.38。
用于测量下肢力量的便携式单轴测力传感器在社区环境中提供可靠的测量结果。