Sridhara Srilekha, Yazdchi Siamac, Alkhatib Jamal T, Berry Anna, Bokhari Ismail A H
Internal Medicine, Banner Heart Hospital, Mesa, USA.
Internal Medicine, Banner University Medical Center, Phoenix, USA.
Cureus. 2019 Feb 11;11(2):e4045. doi: 10.7759/cureus.4045.
Embolic protection devices are used to minimize the risk of stroke by preventing the migration of emboli during carotid artery stenting (CAS). After the successful conclusion of the CAS procedure, these devices are meant to be retrieved. Sometimes retrieval of the filter can be difficult. This difficulty in retrieval can be due to multiple factors such as incomplete stent expansion, stent fracture, vasospasm, and vessel tortuosity causing pseudostenoses. In this case report, an under-expanded proximal carotid stent strut contributed to the filter not being retrievable in spite of maneuvers to retrieve the filter. An innovative approach was used; a coronary guide extension catheter was used to enhance support and balloon-assisted tracking of the extension catheter then permitted advancement of the retrieval device and ultimate retrieval of the filter. This technique to retrieve a carotid filter has not been previously described in the literature.
栓子保护装置用于在颈动脉支架置入术(CAS)期间通过防止栓子迁移来将中风风险降至最低。在CAS手术成功完成后,这些装置应被取出。有时取出过滤器可能会很困难。这种取出困难可能是由于多种因素造成的,如支架扩张不完全、支架断裂、血管痉挛以及导致假性狭窄的血管迂曲。在本病例报告中,尽管采取了取出过滤器的操作,但近端颈动脉支架支柱扩张不足导致过滤器无法取出。采用了一种创新方法;使用冠状动脉导引导管来增强支撑,然后通过球囊辅助追踪该延长导管,从而使取出装置得以推进并最终取出过滤器。这种取出颈动脉过滤器的技术此前在文献中尚未有描述。