种植体间距离和循环脱位对 LOCATOR 和球附着体固位力的影响:一项体外研究。

Effects of interimplant distance and cyclic dislodgement on retention of LOCATOR and ball attachments: An in vitro study.

机构信息

Dental student, Faculty of Dentistry, Shahed University, Tehran, Iran.

Associate Professor, Department of Prosthodontics, Faculty of Dentistry, Shahed University, Tehran, Iran.

出版信息

J Prosthet Dent. 2019 Dec;122(6):550-556. doi: 10.1016/j.prosdent.2018.12.023. Epub 2019 Apr 23.

Abstract

STATEMENT OF PROBLEM

Evaluation of the long-term retentive behavior of overdenture attachments is necessary for successful treatment. Interimplant distance (IID) could affect the retention of these attachments.

PURPOSE

The purpose of this in vitro study was to assess the effects of IID and cyclic dislodgement on the retention of LOCATOR and ball attachments.

MATERIAL AND METHODS

Ball and LOCATOR attachments were connected to corresponding implant analogs. Metal housings were connected to each attachment. Thirty-six pairs of acrylic resin blocks were fabricated for 3 IIDs (19, 23, and 29 mm) and 2 attachments (n=6). Each pair consisted of 2 attachment assemblies parallel to each other with a specified IID. Overall, 1440 dislodgement cycles were applied by using a universal testing machine, and the maximum dislodging force was recorded after 0, 120, 360, 720, and 1440 cycles. Dislodging force data were analyzed by using repeated-measures 3-way analysis of variance, with the number of cycles as the within-subject factor for each specimen. Bonferroni-corrected Student t tests were used to resolve effects that were statistically significant. Moreover, dislodging force data were fit into an exponential decay model to determine the extent and rate of force decay for each attachment and IID studied.

RESULTS

The initial retention of LOCATOR attachments was significantly higher than that of ball attachments with IIDs of 23 and 29 mm, but no significant difference was noted with the 19-mm IID. After 1440 cycles, the retention was statistically similar with that of the 23-mm IID but was significantly higher for ball attachments with IIDs of 19 and 29 mm.

CONCLUSIONS

The IID is one of the factors that affect the retention of ball and LOCATOR attachments. The mean retention of both attachments was sufficient for all 3 IIDs at insertion and after 1440 cycles.

摘要

问题陈述

评估覆盖义齿附着体的长期固位行为对于成功的治疗是必要的。种植体间距离(IID)可能会影响这些附着体的固位。

目的

本体外研究的目的是评估 IID 和循环脱位对 LOCATOR 和球附着体固位的影响。

材料和方法

将球和 LOCATOR 附着体连接到相应的种植体模拟体上。金属外壳连接到每个附着体上。为 3 个 IID(19、23 和 29mm)和 2 个附着体(n=6)制作了 36 对丙烯酸树脂块。每对由 2 个附着体组件组成,彼此平行,具有特定的 IID。总共施加了 1440 次脱位循环,使用万能试验机记录最大脱位力,在 0、120、360、720 和 1440 循环后进行记录。使用重复测量 3 因素方差分析对脱位力数据进行分析,将循环次数作为每个样本的组内因素。使用校正后的学生 t 检验来解决具有统计学意义的影响。此外,将脱位力数据拟合到指数衰减模型中,以确定研究的每个附着体和 IID 的力衰减程度和速率。

结果

LOCATOR 附着体的初始固位力明显高于 23 和 29mm IID 的球附着体,但 19mm IID 无明显差异。经过 1440 次循环后,23mm IID 的固位力统计学上相似,但 19 和 29mm IID 的球附着体的固位力明显更高。

结论

IID 是影响球和 LOCATOR 附着体固位的因素之一。两种附着体的平均固位力在插入时和经过 1440 次循环后对于所有 3 个 IID 均足够。