Bone Mineral Density Precision for Individual and Combined Vertebrae Configurations From Lumbar Spine Dual-Energy X-Ray Absorptiometry Scans.

作者信息

Hind Karen, Oldroyd Brian

机构信息

Department of Sport and Exercise Sciences and the Wolfson Research Institute for Health and Wellbeing, Durham University, Durham, United Kingdom.

School of Sport, Leeds Beckett University, Leeds, United Kingdom.

出版信息

J Clin Densitom. 2020 Oct-Dec;23(4):673-677. doi: 10.1016/j.jocd.2019.04.001. Epub 2019 Apr 5.

Abstract

The accurate interpretation of repeat DXA scan measurements and the understanding of what constitutes a true and meaningful change require knowledge of measurement error (precision) and least significant change. The interpretation of lumbar spine bone mineral density in particular can be confounded by artefacts and as such, the International Society for Clinical Densitometry (ISCD) recommends exclusion of individual vertebrae if they are affected by local structural change or an artefact. The aim of this study was to determine the precision of bone mineral density measures of individual and various configurations of vertebrae from PA lumbar spine scans. The study group comprised of 30 women (age 36.3 ± 6.5 years; height: 165.2 ± 5.7 cm; weight: 67.7 ± 12.6 kg) who each received 2 consecutive anterior-posterior lumbar spine scans (Lunar iDXA, GE Healthcare, Madison, WI), with repositioning. Precision errors varied by individual vertebrae and by different configurations of vertebrae but all were within the ISCD acceptable range of precision. For vertebrae configurations containing at least 2 vertebrae, precision error ranged from 0.005 to 0.008 RMS-SD (0.44%-0.70% CV). Of the individual vertebrae, the lowest precision error was observed at L4, and from the different configurations, for L2L3L4 and L1L2L3L4. In conclusion, this study group demonstrated excellent precision for BMD measurements of individual and various configurations of L1-L4 vertebrae using the GE Lunar iDXA densitometer.

摘要

对重复双能X线吸收法(DXA)扫描测量结果的准确解读以及对何为真实且有意义变化的理解,需要了解测量误差(精密度)和最小有意义变化。特别是腰椎骨密度的解读可能会受到伪影的干扰,因此,国际临床骨密度测量学会(ISCD)建议,如果单个椎体受到局部结构变化或伪影的影响,应将其排除。本研究的目的是确定来自腰椎前后位扫描的单个椎体及不同椎体组合的骨密度测量的精密度。研究组由30名女性组成(年龄36.3±6.5岁;身高:165.2±5.7厘米;体重:67.7±12.6千克),她们每人接受了2次连续的腰椎前后位扫描(Lunar iDXA,GE医疗集团,威斯康星州麦迪逊),且重新进行了定位。精密度误差因单个椎体和椎体的不同组合而异,但均在ISCD可接受的精密度范围内。对于包含至少2个椎体的椎体组合,精密度误差范围为0.005至0.008 RMS-SD(变异系数0.44%-0.70%)。在单个椎体中,L4的精密度误差最低,在不同组合中,L2L3L4和L1L2L3L4的精密度误差最低。总之,本研究组使用GE Lunar iDXA骨密度仪对L1-L4椎体的单个及不同组合进行骨密度测量时,显示出了出色的精密度。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验