Computed tomography (CT) dose optimization is an important issue in radiography because CT is the largest contributor to medical radiation dose and its use is increasing. However, CT dose optimization for pediatric patients could be more challenging than their adult counterparts. The purpose of this literature review was to identify and discuss the current pediatric CT dose saving techniques. Optimized pediatric protocols were also proposed.
A comprehensive literature search was conducted using the Medline, ProQuest Health and Medical Complete, PubMed, ScienceDirect, Scopus, Springer Link, and Web of Science databases and the keywords CT, pediatric, optimization, protocol, and radiation dose to identify articles focusing on pediatric CT dose optimization strategies published between 2004 and 2014.
Seventy-seven articles were identified in the literature search. Strategies for optimizing a range of scan parameters and technical considerations including tube voltage and current, iterative reconstruction, diagnostic reference levels, bowtie filters, scout view, pitch, scan collimation and time, overscanning, and overbeaming for pediatric patients with different ages and body sizes and compositions were discussed. An example of optimized pediatric protocols specific to age and body size for the 64-slice CT scanners was devised. It is expected that this example could provide medical radiation technologists, radiologists, and medical physicists with ideas to optimize their pediatric protocols.
计算机断层扫描(CT)剂量优化是放射成像中的一个重要问题,因为CT是医疗辐射剂量的最大贡献者,且其使用量正在增加。然而,儿科患者的CT剂量优化可能比成人更具挑战性。本文献综述的目的是识别和讨论当前儿科CT剂量节省技术。还提出了优化的儿科扫描方案。
使用Medline、ProQuest健康与医学全文数据库、PubMed、ScienceDirect、Scopus、Springer Link和Web of Science数据库进行全面的文献检索,并使用关键词CT、儿科、优化、方案和辐射剂量,以识别2004年至2014年期间发表的关注儿科CT剂量优化策略的文章。
在文献检索中识别出77篇文章。讨论了针对不同年龄、体型和身体组成的儿科患者优化一系列扫描参数的策略以及技术考量,包括管电压和电流、迭代重建、诊断参考水平、蝴蝶结滤过器、定位像、螺距、扫描准直和时间、过度扫描和过度照射。针对64层CT扫描仪设计了一个根据年龄和体型定制的优化儿科扫描方案示例。预计该示例可为医学放射技师、放射科医生和医学物理师提供优化儿科扫描方案的思路。