Suppr超能文献

用于X波段体内电子顺磁共振剂量测定的牙齿体积和几何形状归一化方法。

A normalization method of the volume and geometry of tooth for X-band in vivo EPR dosimetry.

作者信息

Zou Jierui, Guo Junwang, Ma Lei, Dong Guofu, Cong Jianbo, Liu Ye, Tian Ye, Wu Ke

机构信息

Beijing Institute of Radiation Medicine, Beijing Key Laboratory of Radiation Biology (No. BZ0325), China.

Beijing Institute of Radiation Medicine, Beijing Key Laboratory of Radiation Biology (No. BZ0325), China.

出版信息

Appl Radiat Isot. 2019 Jul;149:123-129. doi: 10.1016/j.apradiso.2019.04.023. Epub 2019 Apr 19.

Abstract

The accuracy of in vivo EPR tooth dosimetry may be influenced by the volume and geometry variations in teeth, especially when there is considerable non-uniform sensitivity distribution in the active detection area of the cavity. To solve this problem, the present research proposed a normalization method specifically for X-band EPR in vivo tooth dosimetry. The volume and geometry of the measured tooth were reconstructed by digital image processing with images of the tooth impression slices, which were obtained by a custom-made impression module. The sensitivity distribution in the active detection area was established based on experiments with a point sample. Consequently, a composite normalization process that could calibrate the evaluated dose effectively was carried out by taking into account the influences not only from tooth volume and geometry but also from the non-uniform distribution of sensitivity. The effect and practicability of the method were evaluated by incisor samples. Results showed that the standard deviation could be reduced a maximum of 54.8% approximately after the composite normalization, an improvement compared to results from solely tooth volume. The correlation coefficient of the dose-response curve could be improved from 0.731 to 0.986. The preliminary method provides an approach potentially useful on site after radiation accidents when dealing with the influence of variations in the tooth volume and geometry for X-band EPR in vivo dose estimations.

摘要

体内电子顺磁共振牙齿剂量测定法的准确性可能会受到牙齿体积和几何形状变化的影响,尤其是当腔体的活性检测区域存在相当大的非均匀灵敏度分布时。为了解决这个问题,本研究提出了一种专门用于X波段体内电子顺磁共振牙齿剂量测定法的归一化方法。通过使用定制的印模模块获取的牙齿印模切片图像,利用数字图像处理技术重建被测牙齿的体积和几何形状。基于点样品实验建立活性检测区域内的灵敏度分布。因此,通过考虑不仅来自牙齿体积和几何形状,还来自灵敏度非均匀分布的影响,进行了一个能够有效校准评估剂量的复合归一化过程。通过切牙样本评估了该方法的效果和实用性。结果表明,复合归一化后标准偏差最多可降低约54.8%,与仅考虑牙齿体积的结果相比有了改进。剂量响应曲线的相关系数可从0.731提高到0.986。该初步方法为辐射事故后现场处理牙齿体积和几何形状变化对X波段体内电子顺磁共振剂量估计的影响提供了一种可能有用的方法。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验