A 模式超声定量人体成分的可靠性和有效性。

Reliability and Validity of A-Mode Ultrasound to Quantify Body Composition.

机构信息

University of Iowa Hospitals and Clinics, Iowa City, IA.

出版信息

J Orthop Trauma. 2019 Sep;33(9):472-477. doi: 10.1097/BOT.0000000000001498.

Abstract

OBJECTIVE

Immobilization and inadequate nutrition after high-energy trauma result in loss of lean muscle mass. Body composition [percent body fat (%BF), fat mass (FM), fat-free mass (FFM)] is traditionally quantified with techniques that require expensive equipment and/or ionizing radiation. The purpose of this study was to assess reliability and validity of amplitude-mode (A-mode) ultrasound (US) to quantify body composition for use in a randomized clinical trial assessing interventions to prevent muscle loss in young individuals after high-energy musculoskeletal trauma.

METHODS

Thirty-one volunteer subjects underwent body composition assessments using A-mode US and air displacement plethysmography (ADP). Independent raters performed 2 serial US measures. The same raters performed body composition assessment in 12 consecutive subjects that sustained acute musculoskeletal trauma indicated for surgery. Test-retest and inter-rater reliability were assessed using intraclass correlation coefficient (ICC). Agreement between US and ADP was assessed with Bland-Altman analysis.

RESULTS

Test-retest and inter-rater reliability was excellent for volunteer subjects, with ICC values (%BF, FM, and FFM) of 0.87, 0.90, and 0.99 (rater 1); 0.80, 0.82, and 0.98 (rater 2). FFM measured by US was strongly correlated with ADP measures (r = 0.9635, P < 0.05). Bland-Altman analysis demonstrated no systematic bias between US and ADP measures of body composition. In trauma subjects, the inter-rater reliability was excellent, with ICCs of 0.96, 0.98, and 0.99 for %BF, FM, and FFM, respectively.

CONCLUSIONS

A-mode US is a relatively low-cost tool that provides reliable estimates of body composition and is a viable alternative for monitoring body composition in young, healthy patients.

摘要

目的

高能创伤后固定和营养不足会导致瘦体重流失。传统上,通过需要昂贵设备和/或电离辐射的技术来定量身体成分[体脂百分比(%BF)、脂肪量(FM)、去脂体重(FFM)]。本研究旨在评估幅度模式(A 模式)超声(US)量化身体成分的可靠性和有效性,以便在一项评估高能肌肉骨骼创伤后年轻个体预防肌肉损失的干预措施的随机临床试验中使用。

方法

31 名志愿者受试者接受幅度模式 US 和空气置换体描记术(ADP)的身体成分评估。独立的评估者进行了 2 次连续 US 测量。同一位评估者对 12 名连续发生急性肌肉骨骼创伤并需要手术的受试者进行了身体成分评估。使用组内相关系数(ICC)评估测试-重测和组内评估者之间的可靠性。使用 Bland-Altman 分析评估 US 与 ADP 之间的一致性。

结果

志愿者受试者的测试-重测和组内评估者之间的可靠性非常好,ICC 值(%BF、FM 和 FFM)分别为 0.87、0.90 和 0.99(评估者 1);0.80、0.82 和 0.98(评估者 2)。US 测量的 FFM 与 ADP 测量的 FFM 呈强相关(r=0.9635,P<0.05)。Bland-Altman 分析表明,US 和 ADP 测量的身体成分之间没有系统偏差。在创伤受试者中,组内评估者之间的可靠性非常好,%BF、FM 和 FFM 的 ICC 分别为 0.96、0.98 和 0.99。

结论

幅度模式 US 是一种相对低成本的工具,可以提供可靠的身体成分估计值,是监测年轻健康患者身体成分的可行替代方法。