新型机械声频仪(Frequencer™)的频率和振幅对黏液粘弹性的影响。

Influence of frequency and amplitude on the mucus viscoelasticity of the novel mechano-acoustic Frequencer™.

机构信息

Department of Chemical Engineering, Polytechnique Montréal, C.P. 6079, Succ. CV Montréal, H3C 3A7, Québec, Canada.

Universitá degli Studi di Milano, Dipartimento di Chimica, via Golgi 19, 20133, Milano, Italy.

出版信息

Respir Med. 2019 Jul;153:52-59. doi: 10.1016/j.rmed.2019.04.011. Epub 2019 May 27.

Abstract

BACKGROUND

Cystic fibrosis affects 1/3200 Caucasians. This genetic disease disturbs the ion and water homeostasis across epithelia, thus rendering mucus more viscous and harder to expel. Conventional treatments rely on the clapping method coupled with postural drainage. Despite the effectiveness of these procedures, they are invasive and enervating.

METHODS

Here we study a new mechano-acoustic treatment device to help patients expectorate excess mucus, the Frequencer™. We test both normal and pathological synthetic mucin solutions (1 % and 4 % by weight) in vitro. We varied the frequency applied (from 20 Hz to 60 Hz) as well as the amplitude (from 50 % to 100 % intensity). Moreover, we assessed the effect of NaCl on mucus rehydration.

RESULTS

A frequency of 40 Hz coupled with a 0.5 gLNaCl solution provokes partial mucus rehydration, regardless of the amplitude selected, as the work of adhesion measurements evidenced.

CONCLUSIONS

Mechanical solicitation is fundamental to help patients affected by cystic fibrosis expectorate mucus. With an operating frequency of 20 Hz to 65 Hz, the Frequencer™ provides a gentler therapy than traditional methods (conventional chest physiotherapy). The Frequencer™ proved to be effective in the homogenization of synthetic mucin solutions in vitro in 20 min and elicited improved effectiveness in a mucin-rich environment.

摘要

背景

囊性纤维化影响每 3200 名白种人中就有 1 人。这种遗传性疾病会扰乱上皮细胞中的离子和水动态平衡,从而使黏液更加粘稠,更难排出。传统的治疗方法依赖于叩击法结合体位引流。尽管这些程序有效,但它们具有侵入性且令人疲惫。

方法

在这里,我们研究了一种新的机械声治疗设备,以帮助患者排出多余的黏液,即 Frequencer™。我们在体外测试了正常和病理合成黏液(重量比为 1%和 4%)。我们改变了施加的频率(从 20 Hz 到 60 Hz)以及幅度(从 50%到 100%强度)。此外,我们评估了 NaCl 对黏液再水合的影响。

结果

40 Hz 的频率结合 0.5 gLNaCl 溶液会引起部分黏液再水合,无论选择的幅度如何,这正如粘附功测量所证明的那样。

结论

机械刺激对于帮助患有囊性纤维化的患者排出黏液至关重要。Frequencer™ 的工作频率为 20 Hz 至 65 Hz,比传统方法(常规胸部物理疗法)更温和。Frequencer™ 已被证明在 20 分钟内可有效均匀化体外合成黏液溶液,并在富含黏液的环境中提高了效果。