J Rehabil Assist Technol Eng. 2018 Nov 18;5:2055668318802558. doi: 10.1177/2055668318802558. eCollection 2018 Jan-Dec.
Ankle dorsiflexor muscle strength is a crucial component of gait.
We describe the development of a simple, hand-held dynamometer to measure the ankle dorsiflexor muscle strength in the sitting position. In addition, we examine its intra- and inter-rater reliability.
Measurements of the peak ankle dorsiflexor muscle strength were obtained by two examiners for 30 ankles of 15 healthy adults at two time points, with a one-day interval between measurements, to determine the inter- and intra-rater reliability. The intraclass correlation coefficients were calculated, and an intraclass correlation coefficient > 0.90 was considered as excellent reliability. A Bland-Altman analysis was used to assess systemic bias. The minimal detectable change in muscle strength was calculated with a confidence level of 95% (MDC).
The reliability of the device was excellent for both intra- (intraclass correlation coefficients [1,3] = 0.94) and inter-rater (intraclass correlation coefficients [2,3] = 0.96) comparisons. No fixed or proportional bias was observed between the two examiners. The MDC was 0.77 N/kg.
Our results indicate the excellent reliability and responsiveness of our device. By obtaining the measurements of dorsiflexor strength while sitting, compensatory motions are suppressed, yielding a more consistent measurement that can be reliably used to detect subtle changes in the ankle dorsiflexor muscle strength.
踝关节背屈肌力量是步态的关键组成部分。
我们描述了一种简单的手持测力计的开发,用于测量坐姿下的踝关节背屈肌力量。此外,我们还检验了其评分者内和评分者间的可靠性。
由两名检查者在两个时间点对15名健康成年人的30个踝关节进行踝关节背屈肌峰值力量测量,测量间隔为一天,以确定评分者间和评分者内的可靠性。计算组内相关系数,组内相关系数>0.90被认为具有极好的可靠性。采用Bland-Altman分析评估系统偏差。计算肌肉力量的最小可检测变化,置信水平为95%(MDC)。
该设备在评分者内(组内相关系数[1,3]=0.94)和评分者间(组内相关系数[2,3]=0.96)比较中可靠性均极佳。两名检查者之间未观察到固定或比例偏差。MDC为0.77N/kg。
我们的结果表明我们的设备具有极好的可靠性和反应性。通过在坐姿下获得背屈肌力量的测量值,可抑制代偿性运动,从而产生更一致的测量结果,可可靠地用于检测踝关节背屈肌力量的细微变化。