The work investigates the effect of proposed novel semi-rigid stabilization device on lumbar segment L2-L3 so as to preserve motion at vertebral level. Here, the biomechanical behavior of intact spine with three instrumented spinal models (semi-rigid stabilization device, rigid implant and dynamic stabilization system NFlex) have been compared under the motion conditions of flexion, extension, bending and twist. Three-dimensional non-linear finite models of intact spine, semi-rigid stabilization device, rigid implant and dynamic stabilization system NFlex were developed in the present study. All the four models were subjected to a combined load of 400 N in axial compression along with 2, 4, 6, 8 and 10 N m as bending moment individually. Dynamic stabilization system NFlex shows the maximum variation in motion and reflects range of motion as 89.7% during lateral bending, 53.4% in flexion, 34.6% in twist and 28.0% in extension with respect to intact spine. However, semi-rigid stabilization device and rigid implant shows the range of motion of 60%, 48.7%, 32% and 21.8% and 60%, 32.3%, 22.3% and 21.7% of intact, respectively, during bending, flexion, twist and extension. Finite element simulation results reveal that semi-rigid stabilization device shows comparatively lower values than dynamic stabilization system NFlex and higher as compared to rigid implant for measured intradiscal pressure and von Mises strain at intervertebral disc-23.
该研究探讨了新型半刚性稳定装置对腰椎L2-L3节段的影响,以保留椎体水平的运动。在此,在屈曲、伸展、侧弯和扭转的运动条件下,对具有三种植入器械的脊柱模型(半刚性稳定装置、刚性植入物和动态稳定系统NFlex)的完整脊柱的生物力学行为进行了比较。本研究建立了完整脊柱、半刚性稳定装置、刚性植入物和动态稳定系统NFlex的三维非线性有限元模型。所有四个模型均单独承受轴向压缩400 N的联合载荷以及2、4、6、8和10 N·m的弯矩。动态稳定系统NFlex在运动方面显示出最大变化,相对于完整脊柱,其在侧弯时的运动范围为89.7%,屈曲时为53.4%,扭转时为34.6%,伸展时为28.0%。然而,半刚性稳定装置和刚性植入物在侧弯、屈曲、扭转和伸展时的运动范围分别为完整脊柱的60%、48.7%、32%和21.8%以及60%、32.3%、22.3%和21.7%。有限元模拟结果表明,对于椎间盘23处测量的椎间盘内压力和冯·米塞斯应变,半刚性稳定装置显示出比动态稳定系统NFlex相对较低的值,而比刚性植入物高。