Anuradha R, Kulkarni D B, Joseph Leena, Kulkarni M S
Radiation Safety Systems Division, Bhabha Atomic Research Centre, Trombay, 400 085, Mumbai, India.
Radiation Safety Systems Division, Bhabha Atomic Research Centre, Trombay, 400 085, Mumbai, India.
Appl Radiat Isot. 2019 Oct;152:52-56. doi: 10.1016/j.apradiso.2019.06.035. Epub 2019 Jun 22.
The use of Rhenium-188 for various therapeutic applications in the field of nuclear medicine has increased in recent years due to its favourable properties like decay scheme, cost effective availability and easy chemistry. Two independent measuring setups were used to standardise Re radioactive solution. The modus operandi of standardisation was 4πβ-γ coincidence technique where the beta detection was done by proportional counting and liquid scintillation counting and the gamma detection was done by using NaI(Tl) detectors. The secondary standard, high pressure ionisation chamber type Centronic IG12, 20A was calibrated with the standardised Re solution and the sensitivity coefficient (pA MBq) was determined. To enhance the accuracy of the commercial radionuclide calibrator and to ensure that patients receive optimum dose of these radiopharmaceuticals, calibration number of the Capintec CRC-15β radionuclide calibrator was also verified. This paper presents the standardisation of Re radioactive solution by primary methods and calibration of BARC secondary standard ionisation chamber system and a Capintec CRC-15β radionuclide calibrator.
近年来,由于铼 - 188具有衰变模式良好、成本效益高且化学性质简单等有利特性,其在核医学领域各种治疗应用中的使用有所增加。使用了两种独立的测量装置来标准化铼放射性溶液。标准化的操作方法是4πβ - γ符合技术,其中β探测通过正比计数和液体闪烁计数进行,γ探测通过使用碘化钠(铊)探测器进行。用标准化的铼溶液对二级标准高压电离室型Centronic IG12、20A进行校准,并确定灵敏度系数(pA/MBq)。为提高商业放射性核素校准仪的准确性,并确保患者接受这些放射性药物的最佳剂量,还对Capintec CRC - 15β放射性核素校准仪的校准系数进行了验证。本文介绍了通过主要方法对铼放射性溶液进行标准化以及对BARC二级标准电离室系统和Capintec CRC - 15β放射性核素校准仪进行校准的过程。