全膝关节置换术中手术平衡与结果指标之间的关系。

Relationship between surgical balancing and outcome measures in total knees.

作者信息

Chu Lauren M, Meere Patrick A, Oh Cheongeun, Walker Peter S

机构信息

New York University Langone Orthopedic Hospital, New York, NY, USA.

Department of Population Health, New York University School of Medicine, New York, NY, USA.

出版信息

Arthroplast Today. 2019 Mar 2;5(2):197-201. doi: 10.1016/j.artd.2019.01.001. eCollection 2019 Jun.

Abstract

BACKGROUND

The purpose of the study was to investigate the accuracy of balancing which could be achieved at total knee surgery and its relation to functional outcomes.

METHODS

During surgery, the forces on the medial and lateral plateaus were measured at 10-15 degrees flexion in 101 patients, using an instrumented tibial trial, with equal forces being targeted. Of the initial 101 cases, 71 cases completed all follow-up visits to 1 year. At each follow-up visit, the function was measured using the Knee Society Scoring System, and varus and valgus laxity angles were measured.

RESULTS

The mean medial/(medial + lateral) compartmental force ratio was 0.52, with a standard deviation of 0.09. The total contact force was 217 Newtons, with a standard deviation of 72 Newtons. No correlations were found between the functional scores and the compartmental force ratio or total contact force. However, the mean varus and valgus laxity angles, 2.8 and 2.3 degrees, respectively, were very close to the angles of normal intact knees.

CONCLUSIONS

The likely reason for the lack of correlation of function was that the large majority of the balancing ratios were within the range 0.4-0.6 but with a wide spread of functional scores typical of total knee study groups. However, the normal varus and valgus angles achieved at follow-up indicated that equal balancing in early flexion was a reasonable surgical target. Using instrumented tibial trials enabled accurate and consistent balancing values to be achieved, as well as normal varus and valgus laxity angles, which may be important in obtaining optimal outcomes.

摘要

背景

本研究的目的是调查全膝关节置换手术中实现平衡的准确性及其与功能结果的关系。

方法

在手术过程中,使用带仪器的胫骨试验,对101例患者在屈膝10 - 15度时测量内侧和外侧平台的受力情况,目标是使两侧受力相等。最初的101例病例中,71例完成了至1年的所有随访。每次随访时,使用膝关节协会评分系统测量功能,并测量内翻和外翻松弛角度。

结果

内侧/(内侧 + 外侧)间室力比的平均值为0.52,标准差为0.09。总接触力为217牛顿,标准差为72牛顿。未发现功能评分与间室力比或总接触力之间存在相关性。然而,平均内翻和外翻松弛角度分别为2.8度和2.3度,与正常完整膝关节的角度非常接近。

结论

功能缺乏相关性的可能原因是,大多数平衡比在0.4 - 0.6范围内,但全膝关节研究组的功能评分差异很大。然而,随访时达到的正常内翻和外翻角度表明,早期屈曲时的平衡是一个合理的手术目标。使用带仪器的胫骨试验能够实现准确且一致的平衡值,以及正常的内翻和外翻松弛角度,这对于获得最佳结果可能很重要。