J Arthroplasty. 2018 Aug;33(8):2502-2505.e12. doi: 10.1016/j.arth.2018.03.067. Epub 2018 Apr 9.
In pursuit to improve soft tissue balancing in total knee arthroplasties (TKAs), a wireless device was introduced to assess femorotibial pressures. The aim of this study was to evaluate the reliability of this device.
After 33 TKAs were balanced by conventional techniques, contact pressures were measured using a wireless sensor 3 times in a row; twice while the examiner was blinded for the result (n = 29); and once while the examiner was able to see the result as visual feedback (n = 32). Femorotibial pressures were measured in the medial and lateral compartments with the knee in 10°, 45°, and 90° of flexion (6 measurements per TKA). Furthermore, both the combined pressure and the difference in pressure between the compartments was calculated throughout the 3 positions (together another 6 measurements per TKA).
The intraclass correlation coefficient between the blind measurements was poor in 2 of the 12 (17%), moderate in 4 of 12 (33%), and good in 6 of 12 (50%) measurements. The intraclass correlation coefficient between the blind and observing measurement was poor in 2 of the 12 (17%), moderate in 6 of 12 (50%), and good in 4 of 12 (33%) measurements. Especially measurements in 10° of flexion are associated with poorer reliability.
The wireless sensor has a moderate to good reliability in 83% of the measurements.
为了提高全膝关节置换术(TKA)中的软组织平衡,引入了一种无线设备来评估股胫压力。本研究的目的是评估该设备的可靠性。
在通过传统技术平衡了 33 例 TKA 后,使用无线传感器连续测量 3 次接触压力;两次测量时检查者对结果不知情(n=29);一次测量时检查者可以看到结果作为视觉反馈(n=32)。在膝关节 10°、45°和 90°的屈曲位,测量股胫内侧和外侧间隙的压力(每例 TKA 有 6 个测量值)。此外,在 3 个位置计算了股胫间隙的综合压力和压力差值(每例 TKA 总共还有 6 个测量值)。
在 12 次测量中,盲测的组内相关系数在 2 次(17%)为差,在 4 次(33%)为中度,在 6 次(50%)为好。盲测和观察测量的组内相关系数在 2 次(17%)为差,在 6 次(50%)为中度,在 4 次(33%)为好。特别是在 10°的屈曲位测量的可靠性较差。
无线传感器在 83%的测量中具有中等至良好的可靠性。