Analytical quality assessment and method comparison of two immunoassays for the measurement of serum cardiac Troponin I in dogs and cats.

作者信息

Irvine Katherine L, McLeish Susan A, Sarvani Elpida, Papasouliotis Kostas

机构信息

Diagnostic Laboratories, Langford Vets, Bristol Veterinary School, University of Bristol, Bristol, UK.

出版信息

Vet Clin Pathol. 2019 Oct;48 Suppl 1:70-77. doi: 10.1111/vcp.12761. Epub 2019 Jul 16.

Abstract

BACKGROUND

Cardiac troponin I (cTnI) is a sensitive and specific biomarker for myocardial injury. Validation of point-of-care (POC) analyzers for cTnI measurement is valuable to the critical care setting, in which rapid results can facilitate prompt diagnoses. An immunoassay for detecting cTnI is available for the POC AIA-360 analyzer (Tosoh Bioscience), but this has not been validated using canine and feline serum.

OBJECTIVES

The objectives were (a) to determine precision, accuracy, and linearity of cTnI measurement using the AIA-360 immunoassay in pooled canine and feline samples, and (b) to compare results for individual canine and feline samples with those obtained using a reference chemiluminescence method (Immulite 1000, Siemens).

METHODS

Intra- and inter-assay repeatability was determined using pooled canine and feline samples, and the coefficient of variation (CV) was calculated for each. Pooled samples were also serially diluted to assess linearity. A modified spike and recovery analysis was performed by mixing pooled samples with different concentrations. Bland-Altman and Deming regression analyses were used to determine bias for individual samples, and the total observed error (TE ) was calculated.

RESULTS

Coefficient of variation values were well within the required maximum of 20%. Linearity was demonstrated over the range of samples tested, and the recovery study showed minimal proportional inaccuracies. Although the correlation between the analyzers was excellent, there was a large mean bias due to relative proportional bias. Total observed error consequently exceeded the total allowable error (TE ).

CONCLUSION

Although, in most respects, the analyzer demonstrated adequate performance, pronounced bias contributed to the large TE , indicating a requirement for analyzer-specific reference intervals.

摘要

背景

心肌肌钙蛋白I(cTnI)是心肌损伤的一种敏感且特异的生物标志物。对用于cTnI检测的即时检验(POC)分析仪进行验证,对于重症监护环境很有价值,因为快速获得结果有助于及时诊断。一种检测cTnI的免疫测定法可用于POC AIA - 360分析仪(东曹生物科学公司),但尚未使用犬猫血清进行验证。

目的

目的是(a)使用AIA - 360免疫测定法测定犬猫混合样本中cTnI检测的精密度、准确性和线性,以及(b)将犬猫个体样本的检测结果与使用参考化学发光法(西门子Immulite 1000)获得的结果进行比较。

方法

使用犬猫混合样本测定批内和批间重复性,并计算每个样本的变异系数(CV)。对混合样本进行系列稀释以评估线性。通过将不同浓度的混合样本混合进行改良加标回收分析。使用Bland - Altman和Deming回归分析确定个体样本的偏差,并计算总观察误差(TE)。

结果

变异系数值远低于所需的最大20%。在所测试的样本范围内显示出线性,回收研究显示比例误差最小。尽管分析仪之间的相关性极佳,但由于相对比例偏差存在较大的平均偏差。因此,总观察误差超过了总允许误差(TE)。

结论

尽管在大多数方面分析仪表现出足够的性能,但明显的偏差导致了较大的TE,这表明需要针对分析仪特定的参考区间。