To evaluate the repeatability of the steady-state pattern electroretinogram (PERG) and full-field flicker electroretinogram (Flicker ERG) protocols, delivered by the office-based Neuro Optic Vision Assessment (NOVA)™ testing platform, in healthy subjects.
Healthy individuals underwent PERG (16° and 24°) and Flicker ERG [fixed luminance (FL) and multi-luminance (ML)] testing protocols. Test-retest repeatability of protocols was calculated using intra-class correlation coefficients (ICC). Reference values of the parameters of the aforementioned tests were also calculated.
The ICCs for the PERG parameters ranged from 0.793 to 0.911 (p < 0.001). The ICCs for the Flicker ERG parameters ranged from 0.968 to 0.994 (p < 0.001). A linear regression analysis was applied to assess the impact of age on ERG responses. Age had a significant impact on all PERG parameters (16° or 24°). The phase response of the FL Flicker ERG significantly decreased with age (β = - 0.837, p ≤ 0.001). The FL Flicker ERG Magnitude was also impacted with a significant quadratic effect of age (β = - 0.0047, p = 0.0004). Similarly, the Phase Area Under the Curve (Phase AUC) of the ML Flicker ERG significantly declined with age (β = - 0.007, p = 0.009), and the impact on the Magnitude AUC was significant as well, with a negative quadratic age effect.
The PERG and Flicker ERG protocols, delivered by an office-based testing platform, were shown to have good-to-excellent test-retest repeatability when tests were performed in the same order and in immediate succession.
评估基于办公室的神经视视觉评估(NOVA)™测试平台所提供的稳态图形视网膜电图(PERG)和全视野闪烁视网膜电图(闪烁ERG)检测方案在健康受试者中的可重复性。
健康个体接受PERG(16°和24°)和闪烁ERG[固定亮度(FL)和多亮度(ML)]检测方案。使用组内相关系数(ICC)计算检测方案的重测可重复性。还计算了上述检测参数的参考值。
PERG参数的ICC范围为0.793至0.911(p<0.001)。闪烁ERG参数的ICC范围为0.968至0.994(p<0.001)。应用线性回归分析评估年龄对ERG反应的影响。年龄对所有PERG参数(16°或24°)均有显著影响。FL闪烁ERG的相位反应随年龄显著降低(β=-0.837,p≤0.001)。FL闪烁ERG幅值也受到年龄的显著二次效应影响(β=-0.0047,p=0.0004)。同样,ML闪烁ERG的曲线下相位面积(相位AUC)随年龄显著下降(β=-0.007,p=0.009),对幅值AUC的影响也显著,存在负的年龄二次效应。
当以相同顺序并紧接着进行检测时,基于办公室的测试平台所提供的PERG和闪烁ERG检测方案显示出良好至优秀的重测可重复性。