双极半关节置换装置线性磨损的影像学评估

Radiographic evaluation of linear wear of bipolar hemiarthroplasty devices .

作者信息

Kaku Nobuhiro, Noda Shouhei, Tabata Tomonori, Tagomori Hiroaki, Tsumura Hiroshi

机构信息

Department of Orthopaedic Surgery, Faculty of Medicine, Oita University, 1-1 Idaigaoka Hazama-machi, Yufu City, Oita, Japan.

出版信息

J Clin Orthop Trauma. 2019 Jul-Aug;10(4):639-644. doi: 10.1016/j.jcot.2018.07.016. Epub 2018 Jul 20.

Abstract

BACKGROUND

The indications for bipolar hemiarthroplasty (BHA) have been narrowing as those for total hip arthroplasty (THA) have been expanding in cases of femoral neck fracture and initial stages of osteonecrosis of femoral head (ONFH). It is difficult to measure wear after BHA because the equipment used has a dual bearing system with an inner head inside a metal outer cup. The present study aimed to (1) use a roentgenograph to measure linear wear after BHA with different acetabular conditions around the bipolar cup, and (2) compare the linear wear of polyethylene between integral bipolar cup (IBC) and Centrax prostheses.

METHODS

From among patients who had undergone BHA in our department after 1996, we analyzed 48 joints with osteoarthritis (OA) and without acetabular cartilage, as well as 25 joints with ONFH of stage 3 or less, in which the acetabular cartilage remains. Two types of bipolar cup prostheses were used: the IBC, which was gamma-sterilized in air using 2-Mrad irradiation, and the Centrax, which was gamma-sterilized in nitrogen using 3-Mrad irradiation. To image the inner head , we used high-pressure X-ray photography; we measured linear wear of the polyethylene with software using Martell Method 1.

RESULTS

Comparing mean annual linear wear between the OA and ONFH groups, using IBC prostheses in both groups, there was a significant difference (0.213 mm vs. 0.096 mm, respectively;  = ). There was a significant difference between the Centrax and IBC prostheses in OA patients in terms of linear wear (0.04 mm vs. 0.213 mm;  = 0.0181).

CONCLUSION

The linear wear of polyethylene in BHA implants can be affected by the material used to manufacture the bipolar cup. Such implants should only be used for appropriate indications.

摘要

背景

在股骨颈骨折和股骨头坏死(ONFH)初期病例中,随着全髋关节置换术(THA)的适应证不断扩大,双极半髋关节置换术(BHA)的适应证却在不断缩小。由于BHA所使用的设备具有双轴承系统,金属外杯内有一个内股骨头,因此很难测量其磨损情况。本研究旨在:(1)使用X线片测量双极杯周围不同髋臼条件下BHA术后的线性磨损;(2)比较一体化双极杯(IBC)和Centrax假体之间聚乙烯的线性磨损。

方法

从1996年以后在我科接受BHA手术的患者中,我们分析了48例患骨关节炎(OA)且无髋臼软骨的关节,以及25例髋臼软骨尚存的3期及以下ONFH关节。使用了两种双极杯假体:IBC,在空气中用2兆拉德辐射进行γ灭菌;Centrax,在氮气中用3兆拉德辐射进行γ灭菌。为了对内股骨头进行成像,我们使用了高压X线摄影;我们使用Martell方法1软件测量聚乙烯的线性磨损。

结果

在两组均使用IBC假体的情况下,比较OA组和ONFH组的平均年线性磨损,存在显著差异(分别为0.213毫米和0.096毫米;P = )。在OA患者中,Centrax和IBC假体在线性磨损方面存在显著差异(0.04毫米对0.213毫米;P = 0.0181)。

结论

BHA植入物中聚乙烯的线性磨损可能会受到制造双极杯所用材料的影响。此类植入物应仅用于适当的适应证。