MoistureMeterD 紧凑型设备和凹陷试验评估乳腺癌相关淋巴水肿患者局部组织水分的可靠性。

Reliability of the MoistureMeterD Compact Device and the Pitting Test to Evaluate Local Tissue Water in Subjects with Breast Cancer-Related Lymphedema.

机构信息

Department of Rehabilitation Sciences, KU Leuven-University of Leuven, Leuven, Belgium.

Department of Rehabilitation Sciences and Physiotherapy, MOVANT, University of Antwerp, Antwerp, Belgium.

出版信息

Lymphat Res Biol. 2020 Apr;18(2):116-128. doi: 10.1089/lrb.2019.0013. Epub 2019 Jul 19.

Abstract

Local tissue water in patients with breast cancer-related lymphedema (BCRL) can be assessed by measurement of the tissue dielectric constant using the MoistureMeterD Compact (MMDC) device, or by performing the pitting test. Although these assessment methods are commonly used in clinical practice, literature shows a lack of research on their clinimetric properties. Therefore, the aim of this study was to investigate reliability of both methods, in assessing the upper limb in BCRL. Thirty women with BCRL were enrolled. Local tissue water was evaluated at nine reference points on the upper limb and trunk, using both methods. To determine intra- and inter-rater reliability of the MMDC device (using the absolute percentages of water content [PWC%] and interarm PWC% ratios based on single and multiple measures), intraclass correlation coefficients (ICCs), and standard errors of the measurement were calculated. To determine intra- and inter-rater agreement of the pitting test, Cohen's kappa coefficients were calculated as well as percentages of agreement. MMDC measurements yielded moderate to very strong intra- (ICC 0.648-0.947) and inter-rater (ICC 0.606-0.941) reliability, depending on the measurement location on the edematous limb. The pitting test showed a very strong intrarater agreement at nearly all defined points, but a weak inter-rater agreement, especially at the medial elbow and the breast. This study supports the MMDC device and pitting test as being useful tools in the clinical evaluation of BCRL. However, further research into the concurrent validity of both tools is warranted.

摘要

采用 MoistureMeterD Compact (MMDC) 设备或进行凹陷试验来测量组织介电常数,可评估患有乳腺癌相关淋巴水肿 (BCRL) 患者的局部组织含水量。尽管这些评估方法在临床实践中很常用,但文献表明,关于这些评估方法的计量学特性的研究很少。因此,本研究旨在调查这两种方法评估 BCRL 上肢的可靠性。

纳入了 30 名患有 BCRL 的女性。使用这两种方法在上肢和躯干的 9 个参考点评估局部组织含水量。为了确定 MMDC 设备的内部和内部评估者可靠性(使用绝对含水量百分比 [PWC%] 和基于单次和多次测量的双臂间 PWC% 比值),计算了组内相关系数 (ICC) 和测量标准误差。为了确定凹陷试验的内部和外部评估者一致性,还计算了 Cohen's kappa 系数和一致性百分比。

根据水肿肢体上的测量位置,MMDC 测量结果显示出中等至非常强的内部(ICC 0.648-0.947)和外部(ICC 0.606-0.941)评估者可靠性。凹陷试验在几乎所有定义的点都显示出很强的内部评估者一致性,但外部评估者一致性较弱,尤其是在肘内侧和乳房处。

本研究支持 MMDC 设备和凹陷试验作为评估 BCRL 的有用工具。然而,需要进一步研究这两种工具的同时有效性。