Schou-Pedersen Anne Marie V, Schemeth Dieter, Lykkesfeldt Jens
Section for Experimental Animal Models, Department of Veterinary and Animal Sciences, University of Copenhagen, Ridebanevej 9, 1870 Frederiksberg C, Denmark.
Section for Environmental Chemistry and Physics, Department of Plant and Environmental Sciences, Thorvaldsensvej 40, 1870 Frederiksberg C, Denmark.
Antioxidants (Basel). 2019 Jul 30;8(8):253. doi: 10.3390/antiox8080253.
Coenzyme Q (Q) plays an important role in mammals for energy production in the mitochondria, and as a potent antioxidant. Oxidation ratio (% oxidized in relation to total Q) has been proposed as an important biomarker. A sensitive and reproducible HPLC-ECD method was developed for determination of reduced and oxidized Q in canine plasma and heart tissue. Chromatographic separation was achieved in 10 min using a Waters Nova-pak C column and a mobile phase with lithium perchlorate in ethanol/methanol/2-propanol. The validation showed satisfying results. Excellent linear correlation was found ( > 0.9997), intra- and inter-day precisions were below 6.5% ( = 5) and recoveries were between 89 and 109% ( = 5). Sensitivity stated as Lower Limit of Quantification (LLOQ) was 10 nM. Acceptable stability of both extracted and un-extracted samples was observed. The plasma concentration range of total Q was found to be between 0.64 and 1.24 µg/mL. Comparison with a developed LC-MS/MS method showed a correlation of = 0.85 for reduced Q and = 0.60 for oxidized Q ( = 17). However, average results were around 30% lower for ubiquinol using the LC-MS/MS method as compared with the HPLC-ECD analysis. The two methods are therefore not considered to be interchangeable.
辅酶Q(Q)在哺乳动物线粒体能量产生过程中发挥着重要作用,并且是一种强效抗氧化剂。氧化率(相对于总Q的氧化百分比)已被提议作为一种重要的生物标志物。我们开发了一种灵敏且可重复的高效液相色谱 - 电化学检测法(HPLC - ECD),用于测定犬血浆和心脏组织中还原型和氧化型Q。使用沃特世Nova - pak C柱以及含有高氯酸锂的乙醇/甲醇/2 - 丙醇流动相,在10分钟内实现了色谱分离。验证结果令人满意。发现具有出色的线性相关性(> 0.9997),日内和日间精密度均低于6.5%(n = 5),回收率在89%至109%之间(n = 5)。以定量下限(LLOQ)表示的灵敏度为10 nM。观察到提取和未提取样品均具有可接受的稳定性。发现总Q的血浆浓度范围在0.64至1.24 µg/mL之间。与已开发的液相色谱 - 串联质谱法(LC - MS/MS)比较,还原型Q的相关性为r = 0.85,氧化型Q的相关性为r =
0.60(n = 17)。然而,与HPLC - ECD分析相比,使用LC - MS/MS方法测定的泛醇平均结果低约30%。因此,这两种方法不被认为是可互换的。