Fujitani R M, White R A, Kopchok G E, Vlasak J, Marcus C S, White G H
Department of Surgery, Harbor--University of California, Los Angeles Medical Center 90509.
J Vasc Surg. 1988 Sep;8(3):274-9.
The thrombogenicity of argon laser-assisted vascular anastomoses (LAVAs) was compared with that of sutured vascular anastomoses (SVAs) by measurement of platelet aggregation at the site of repair in a canine model. Sequential 1 cm longitudinal carotid and femoral arteriotomies (n = 80) or jugular and femoral phlebotomies (n = 80) were performed, with each vessel having two tandem, randomly positioned arteriotomies or phlebotomies separated by a 4 cm length of intact vessel. One incision was repaired by SVA with continuous 6-0 polypropylene sutures and the other by argon LAVA. For the laser fusions, argon laser energy was applied to the adventitial surface of the vessel with a 300 micron fiberoptic probe with 0.5 W power, 1100 joules per square centimeter energy fluence, and 150 second exposure per 1 cm length. The arterial and venous segments of SVAs and LAVAs and an equivalent length of normal vessel were harvested at 48 hours (n = 16, 16, 16), 2 weeks (n = 12, 12, 12), and 4 weeks (n = 12, 12, 12). Autologous indium 111 oxine-labeled platelets were injected intravenously 48 hours before removal of the vascular repairs and the radioactivity of the specimens was determined on removal with a NaI (T1) well-type scintillation counter. Anastomotic platelet adherence index (APAI) was calculated as the ratio of emissions of SVA or LAVA to normal reference vessel.(ABSTRACT TRUNCATED AT 250 WORDS)