Bihari Shailesh, Galluccio Steven
Clin Lab. 2019 Aug 1;65(8). doi: 10.7754/Clin.Lab.2019.190121.
Accurate chloride measurement is important in critically ill patients.
Chloride concentration measured simultaneously between the central laboratory (indirect ion-selective electrode) and blood gas analysis (direct ion-selective electrode) were compared.
We report a discrepancy with chloride measurement between the central laboratory and blood gas analysis at low bicarbonate levels.
Caution should be applied while interpreting the chloride concentration when indirect ion-selective electrode methodology is used, especially in the setting of low serum bicarbonate levels.