Matched Cohort Comparison Indicates Superiority of Precurved Electrode Arrays.

机构信息

Department of Hearing and Speech Sciences.

Department of Otolaryngology-Head and Neck Surgery, Vanderbilt University Medical Center, Nashville, Tennessee.

出版信息

Otol Neurotol. 2019 Oct;40(9):1160-1166. doi: 10.1097/MAO.0000000000002366.

Abstract

OBJECTIVE

Characterize differences in adult cochlear implant outcomes and programming parameters for a straight (CI422/522) and a precurved (CI532) electrode array.

SETTING

Cochlear implant (CI) program at a tertiary otologic center.

PATIENTS

Fifty-eight adults were included in the study; 29 were implanted with CI422 or CI522 and 29 were implanted with CI532. Each CI532 recipient was matched to a CI422/522 recipient in terms of age and preoperative hearing thresholds for comparison purposes.

MAIN OUTCOME MEASURES

Consonant-Nucleus-Consonant (CNC) words, AzBio sentences, residual audiometric thresholds, and Speech Spatial Qualities (SSQ) questionnaire collected 6 months postoperatively were used to characterize outcomes. Pulse duration, maxima, impedances, and overall charge measurements were used to characterize programming parameters.

RESULTS

Postoperative unaided low frequency pure-tone average (LFPTA) was significantly better for the CI532 group. CNC scores were significantly better for the CI532 group. Impedances and pulse duration were significantly lower for the CI532 group, but there was no difference in overall charge between the groups.

CONCLUSION

The CI532 group showed either similar or statistically superior results on all measures when compared with the CI422/522 suggesting that the CI532 electrode may be an advantageous substitute for the CI522.

摘要

目的

描述直形(CI422/522)和预弯形(CI532)电极阵列在成人人工耳蜗植入中的差异和编程参数。

环境

在一家三级耳科中心的人工耳蜗(CI)项目。

患者

58 名成年人被纳入研究;29 名植入 CI422 或 CI522,29 名植入 CI532。为了比较目的,每位 CI532 受者与 CI422/522 受者在年龄和术前听力阈值方面相匹配。

主要观察指标

术后 6 个月收集的辅音-核-辅音(CNC)词、AzBio 句子、残余听阈和言语空间质量(SSQ)问卷用于描述结果。使用脉冲持续时间、最大值、阻抗和总电荷测量来描述编程参数。

结果

CI532 组术后未助听低频纯音平均(LFPTA)明显更好。CI532 组的 CNC 评分明显更好。CI532 组的阻抗和脉冲持续时间明显较低,但两组的总电荷无差异。

结论

与 CI422/522 相比,CI532 组在所有测量指标上均表现出相似或统计学上更优的结果,这表明 CI532 电极可能是 CI522 的有利替代物。

相似文献

引用本文的文献

3
The Effect of Cochlear Implant Electrode Array Type on Hearing Preservation.人工耳蜗电极阵列类型对听力保留的影响。
J Otolaryngol Head Neck Surg. 2025 Jan-Dec;54:19160216251316217. doi: 10.1177/19160216251316217. Epub 2025 Mar 24.
6
Cochlear implant electrode design for safe and effective treatment.用于安全有效治疗的人工耳蜗电极设计
Front Neurol. 2024 May 2;15:1348439. doi: 10.3389/fneur.2024.1348439. eCollection 2024.

本文引用的文献

4
Clinical investigation of the Nucleus Slim Modiolar Electrode.Nucleus Slim蜗轴电极的临床研究。
Audiol Neurootol. 2017;22(3):169-179. doi: 10.1159/000480345. Epub 2017 Oct 24.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验