Williams Brendan A, Gil Jorge, Farmer Kevin W
Department of Orthopaedic Surgery, Children's Hospital of Philadelphia, Philadelphia, PA, USA.
Department of Orthopaedics & Rehabilitation, University of Florida, Gainesville, FL, USA.
Case Rep Orthop. 2019 Aug 19;2019:1061385. doi: 10.1155/2019/1061385. eCollection 2019.
A unique case of late migration of an adjustable-loop femoral fixation button utilized during anterior cruciate ligament (ACL) reconstruction is presented. Imaging and physical examinations during the patient's postoperative course were unremarkable for graft or hardware failure. Two years postoperatively, symptomatic hardware migration occurred requiring arthroscopic removal. To our knowledge, this is the first reported case of late migration of an adjustable-loop femoral fixation button in ACL reconstruction. This case highlights that late loosening and migration of adjustable-loop femoral fixation devices in ACL reconstruction can occur despite demonstrated postoperative radiographic and clinical stability. Surgeons utilizing this fixation device should be aware of this potential complication to avoid delayed recognition and patient morbidity.
本文介绍了一例前交叉韧带(ACL)重建术中使用的可调环股骨固定纽扣延迟移位的独特病例。患者术后过程中的影像学检查和体格检查未发现移植物或硬件故障。术后两年,出现有症状的硬件移位,需要关节镜下取出。据我们所知,这是第一例报道的ACL重建中可调环股骨固定纽扣延迟移位的病例。该病例突出表明,尽管术后影像学和临床检查显示稳定,但ACL重建中可调环股骨固定装置仍可能发生晚期松动和移位。使用这种固定装置的外科医生应意识到这种潜在并发症,以避免延误诊断和患者发病。