Department of Biochemistry and Hormonology, CHU de Montpellier, France; PhyMedExp, University of Montpellier, INSERM, CNRS, France.
Department of Biochemistry and Hormonology, CHU de Montpellier, France.
Clin Biochem. 2019 Dec;74:47-53. doi: 10.1016/j.clinbiochem.2019.09.005. Epub 2019 Sep 15.
Blood gas analyzers are frequently installed as point of care devices and thus allow rapid decision making. Few data are available regarding analytical performance of large sets of BGA. We aimed at evaluating 22 ABL 90 Flex Plus analyzers intended to be deployed. The evaluation was performed at the device level and at the entire set level to characterize the quality of measurements but also to ensure consistency across the devices deployed in the hospital.
Imprecision and total error were assessed for pH, pCO2, pO2, sodium, potassium, ionized calcium, glucose, lactate and oximetry parameters. Imprecision at the hospital level including between device variability was also evaluated. One of the two analyzers used in the central laboratory was correlated with a GEM Premier 4000 and a Cobas b221 analyzers. Thereafter, we tested sequentially the 20 instruments intended to be deployed in care service in comparison with the reference device.
Heterogeneity of analytical performance across the different analyzers was low, allowing to consider the whole set as a unique analyzer. The total error was in line with performance goals. Analytical performance of the analyzers was found suitable for use in clinical practice.
Our study is an example of the qualification of a set of point and underscores 1)The need for a unified qualification scheme when multiple analyzers are deployed simultaneously 2) analytical performance goals compatible with clinical use and the state of the art for all parameters.
血气分析仪常被用作床边设备,从而可以快速做出决策。目前有关大量血气分析仪分析性能的数据较少。我们旨在评估 22 台 ABL 90 Flex Plus 分析仪,这些仪器将被部署。评估在仪器级别和整个仪器组级别进行,以评估测量质量,同时确保医院中部署的所有仪器的一致性。
评估 pH 值、pCO2、pO2、钠、钾、离子钙、葡萄糖、乳酸和血氧饱和度参数的精密度和总误差。还评估了包括仪器间变异性在内的医院级别的精密度。中心实验室使用的两台分析仪中的一台与 GEM Premier 4000 和 Cobas b221 分析仪相关联。此后,我们对计划部署在护理服务中的 20 台仪器进行了顺序测试,与参考仪器进行了比较。
不同仪器之间的分析性能存在低异质性,允许将整个仪器组视为一个独特的仪器。总误差符合性能目标。分析仪的分析性能适合临床实践使用。
我们的研究是一套仪器资格认证的示例,突出了以下几点:1)当同时部署多台分析仪时,需要统一的资格认证方案;2)与临床使用和所有参数的最新技术兼容的分析性能目标。