Lippi Giuseppe, Pighi Laura, Salvagno Gian Luca, Tiziani Elena, Castellini Maria Elena, Ferraro Roberta, Henry Brandon M
Section of Clinical Biochemistry and School of Medicine, University of Verona, Verona, Italy.
Clinical Laboratory, Division of Nephrology and Hypertension, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA.
Adv Lab Med. 2024 May 7;5(3):301-306. doi: 10.1515/almed-2024-0043. eCollection 2024 Sep.
The purpose of this study was to investigate the effects of contamination of venous blood with a lipid-containing solution on parameters measured by a modern blood gas analyzer.
We collected venous blood from 17 healthcare workers (46 ± 11 years; 53 % women) into three blood gas syringes containing 0 , 5 and 10 % lipid-containing solution. Blood gas analysis was performed within 15 min from sample collection on GEM Premier 5000, while triglycerides and serum indices were assays on Roche COBAS C702.
Triglycerides concentration increased from 1.0 ± 0.3 mmol/L in the uncontaminated blood gas syringe, to 39.4 ± 7.8 and 65.3 ± 14.4 mmol/L (both p<0.001) in syringes with 5 and 10 % final lipid contamination. The lipemic and hemolysis indices increased accordingly. Statistically significant variation was noted for all analytes except hematocrit and COHb in the syringe with 5 % lipids, while only COHb did not vary in the syringe with 10 % lipids. Significant increases were observed from 5 % lipid contamination for pO, SO and lactate, while the values of pH, pCO, sodium, potassium, chloride, ionized calcium, glucose, hematocrit (10 % contamination), hemoglobin and MetHB decreased. All these changes except lactate and CoHb exceeded their relative performance specifications.
Artifactual hyperlipidemia caused by contamination with exogenous lipids can have a clinically significant impact on blood gas analysis. Manufacturers of blood gas analyzers must be persuaded to develop new instruments equipped with serum indices.
本研究旨在调查含脂溶液污染静脉血对现代血气分析仪所测参数的影响。
我们从17名医护人员(46±11岁;53%为女性)采集静脉血,分别注入三个装有0%、5%和10%含脂溶液的血气注射器中。采集样本后15分钟内,在GEM Premier 5000上进行血气分析,同时在罗氏COBAS C702上检测甘油三酯和血清指标。
甘油三酯浓度从未受污染的血气注射器中的1.0±0.3毫摩尔/升,增加到最终脂质污染为5%和10%的注射器中的39.4±7.8和65.3±14.4毫摩尔/升(均p<0.001)。脂血指数和溶血指数相应增加。在含5%脂质的注射器中,除血细胞比容和碳氧血红蛋白外,所有分析物均有统计学显著变化,而在含10%脂质的注射器中只有碳氧血红蛋白没有变化。从5%脂质污染开始,氧分压、氧饱和度和乳酸有显著增加,而pH值、二氧化碳分压、钠、钾、氯、离子钙、葡萄糖、血细胞比容(10%污染)、血红蛋白和高铁血红蛋白的值降低。除乳酸和碳氧血红蛋白外,所有这些变化均超出其相对性能规格。
外源性脂质污染导致的人为高脂血症可对血气分析产生临床显著影响。必须说服血气分析仪制造商开发配备血清指标的新仪器。