Suppr超能文献

[角膜的生物力学特性与眼压测量]

[Biomechanical properties of the cornea and tonometry measurements].

作者信息

Bubnova I A, Asatryan S V

机构信息

Research Institute of Eye Diseases, 11A Rossolimo St., Moscow, Russian Federation, 119021.

出版信息

Vestn Oftalmol. 2019;135(4):27-32. doi: 10.17116/oftalma201913504127.

Abstract

PURPOSE

To study peculiarities of intraocular pressure (IOP) measurement by different methods in patients with changes in corneal biomechanical properties.

MATERIAL AND METHODS

The study enrolled 154 patients (290 eyes) divided into 3 groups according to the condition of their corneas in terms of biomechanical properties. Group I consisted of 84 patients (168 eyes) with conditionally 'normal' cornea and corneal thickness ranging from 470 to 660 μm. Group II included 44 patients (70 eyes) with keratokonus. Group III consisted of 26 patients (52 eyes) after LASIK. IOP pressure was measured in each patient by pneumotonometry, Maklakov tonometry, and dynamic bidirectional pneumo-applanation of the cornea. Biomechanical properties were evaluated with dynamic pneumo-impression method.

RESULTS

The study revealed that in patients of groups I and III, corneal biomechanics affected indices of corneal compensated IOP and Maklakov's IOP to a lesser extent, as opposed to measurements of Goldman tonometry and pneumotonometry. In group II, Maklakov's tonometry had less IOP measurement inaccuracies in comparison to other evaluated methods due to larger applanation surface.

CONCLUSION

Changes in biomechanical properties of the cornea can complicate interpretation of the results of different methods of IOP measurement. Accurate determination of IOP requires consideration of biomechanical properties of the cornea, as well as anamnesis data on past corneal diseases or keratorefractive surgeries.

摘要

目的

研究角膜生物力学特性改变的患者采用不同方法测量眼压(IOP)的特点。

材料与方法

该研究纳入了154例患者(290只眼),根据角膜生物力学特性状况分为3组。第一组由84例患者(168只眼)组成,角膜条件“正常”,角膜厚度在470至660μm之间。第二组包括44例圆锥角膜患者(70只眼)。第三组由26例接受准分子激光原位角膜磨镶术(LASIK)后的患者(52只眼)组成。采用空气眼压计、马克拉科夫眼压计和角膜动态双向气动压平法测量每位患者的眼压。采用动态气动压痕法评估生物力学特性。

结果

研究表明,与戈德曼眼压计和空气眼压计的测量结果相比,在第一组和第三组患者中,角膜生物力学对角膜补偿眼压和马克拉科夫眼压指标的影响较小。在第二组中,由于压平面积较大,与其他评估方法相比,马克拉科夫眼压计测量眼压的误差较小。

结论

角膜生物力学特性的改变会使不同眼压测量方法的结果解读变得复杂。准确测定眼压需要考虑角膜的生物力学特性以及既往角膜疾病或角膜屈光手术的病史资料。