Schäfer Stefan B, Papst Sabine, Fiebich Martin, Rudolph Claudia, de Laffolie Jan, Krombach Gabriele A
Department of Diagnostic and Interventional Radiology, University Hospital Giessen, Justus-Liebig University, Klinikstrasse 33, 35392, Giessen, Germany.
Department of Pediatric Radiology, University Hospital Giessen, Justus-Liebig University, 35392, Giessen, Germany.
Pediatr Radiol. 2020 Jan;50(1):28-37. doi: 10.1007/s00247-019-04522-1. Epub 2019 Oct 4.
The acquisition of chest radiographs in neonates is of critical importance in diagnostics because of the risk of respiratory distress syndrome and pneumothorax in preterm infants.
To achieve a dose reduction while preserving a diagnostic image quality for chest radiographs of neonates.
All radiographs, generated on a fully digital X-ray unit by using a neonatal chest phantom, were evaluated under variation of the tube voltage (40-70 kV) and mAs levels (1-10.2 mAs) with and without an additional 0.1-mm copper (Cu) filtration. Noise, contrast and contrast-to-noise ratio for bronchus, heart, lungs and vessels were determined. Visual assessment of the image quality was carried out by three radiologists using a Likert scale. To evaluate a maximally possible dose reduction, the dose of the radiographs with still acceptable image quality at a minimal dose was compared to the dose of the radiographs with the standard settings used in clinical routine.
The noise showed decreasing values with increasing dose, while the contrast values were increased. For the contrast-to-noise ratio, a digressive course of the values as a function of the tube voltage was found. The visual evaluation of image quality showed the best evaluation of the structures at the lowest possible dose in the settings (44 kV, 3.36 mAs) with copper filtration and in the settings (44 kV, 1.56 mAs) without copper filtration. A maximum dose reduction from 8.29 μSv to 2.21 μSv (about 73%) was obtained.
A dose reduction while preserving diagnostic image quality in a digital X-ray system is generally possible by reducing the tube voltage and simultaneous adaptation of the mAs settings.
由于早产儿存在呼吸窘迫综合征和气胸的风险,新生儿胸部X光片的获取在诊断中至关重要。
在保持新生儿胸部X光片诊断图像质量的同时实现剂量降低。
使用新生儿胸部模型在全数字X射线设备上生成的所有X光片,在管电压(40 - 70 kV)和毫安秒(mAs)水平(1 - 10.2 mAs)变化的情况下,分别在有和没有额外0.1毫米铜(Cu)滤过的条件下进行评估。测定支气管、心脏、肺部和血管的噪声、对比度及对比度噪声比。由三位放射科医生使用李克特量表对图像质量进行视觉评估。为评估最大可能的剂量降低,将在最小剂量下仍具有可接受图像质量的X光片剂量与临床常规使用的标准设置下的X光片剂量进行比较。
噪声值随剂量增加而降低,而对比度值增加。对于对比度噪声比,发现其值随管电压呈递减趋势。图像质量的视觉评估显示,在有铜滤过的设置(44 kV,3.36 mAs)和没有铜滤过的设置(44 kV,1.56 mAs)中,在尽可能低的剂量下对结构的评估最佳。最大剂量从8.29 μSv降低到2.21 μSv(约73%)。
在数字X射线系统中,通过降低管电压并同时调整mAs设置,通常可以在保持诊断图像质量的同时降低剂量。