Department of Bioengineering, Imperial College London, London SW7 2AZ, UK.
J Biomech Eng. 2020 Apr 1;142(4). doi: 10.1115/1.4045115.
The accurate measurement of full six degrees-of-freedom (6DOFs) knee joint kinematics is prohibited by soft tissue artifact (STA), which remains the greatest source of error. The purpose of this study was to present and assess a new femoral clamp to reduce STA at the thigh. It was hypothesized that the device can preserve the natural knee joint kinematics pattern and outperform a conventional marker mounted rigid cluster during gait. Six healthy subjects were asked to walk barefoot on level ground with a cluster marker set (cluster gait) followed by a cluster-clamp-merged marker set (clamp gait) and their kinematics was measured using the cluster method in cluster gait and the cluster and clamp methods simultaneously in clamp gait. Two operators performed the gait measurement. A 6DOFs knee joint model was developed to enable comparison with the gold standard knee joint kinematics measured using a dual fluoroscopic imaging technique. One-dimensional (1D) paired t-tests were used to compare the knee joint kinematics waveforms between cluster gait and clamp gait. The accuracy was assessed in terms of the root-mean-square error (RMSE), coefficient of determination, and Bland-Altman plots. Interoperator reliability was assessed using the intraclass correlation coefficient (ICC). The result showed that the femoral clamp did not change the walking speed and knee joint kinematics waveforms. Additionally, clamp gait reduced the rotation and translation errors in the transverse plane and improved the interoperator reliability when compared to the rigid cluster method, suggesting a more accurate and reliable measurement of knee joint kinematics.
软组织伪影(STA)限制了全六自由度(6DOFs)膝关节运动学的精确测量,这仍然是最大的误差源。本研究的目的是提出并评估一种新的股骨夹具,以减少大腿处的 STA。假设该装置可以保持膝关节运动学的自然模式,并在步态中优于传统的刚性标记簇。
要求 6 名健康受试者赤脚在平地行走,分别采用标记簇(簇步态)和标记簇-夹具融合标记集(夹具步态),并用标记簇法在簇步态中、同时用标记簇和夹具法在夹具步态中测量其运动学。由两名操作人员进行步态测量。开发了一个 6DOFs 膝关节模型,以与使用双荧光透视成像技术测量的金标准膝关节运动学进行比较。采用一维(1D)配对 t 检验比较簇步态和夹具步态的膝关节运动学波形。以均方根误差(RMSE)、决定系数和 Bland-Altman 图评估准确性。采用组内相关系数(ICC)评估操作者间的可靠性。结果表明,股骨夹具不会改变行走速度和膝关节运动学波形。此外,与刚性标记簇方法相比,夹具步态减少了横向平面的旋转和平移误差,并提高了操作者间的可靠性,这表明膝关节运动学的测量更准确和可靠。