Centre for Heart Valve Innovation, St. Paul's Hospital, University of British Columbia, Vancouver, BC, Canada.
EuroIntervention. 2020 Feb 7;15(15):e1335-e1342. doi: 10.4244/EIJ-D-19-00782.
We aimed to assess the impact of implant depth on hydrodynamic function following valve-in-valve (VIV) intervention using the ALLEGRA transcatheter heart valve (THV) in three different surgical valve designs.
Multiple implantation depths (+2 mm, -2 mm and -6 mm) were tested using a 23 mm ALLEGRA THV for VIV intervention in 19 mm, 21 mm, 23 mm, and 25 mm Epic, Mitroflow and Magna Ease bioprosthetic valves. Multimodality imaging and hydrodynamic evaluation was performed at each implantation depth. The 23 mm ALLEGRA valve had gradients <20 mmHg in the Mitroflow and Epic valves sized ≥21 mm, and in all sizes of the Magna Ease valve. Gradients did not increase significantly at lower implantation depths. The 19 mm Epic (+2 mm: 20.1±0.6 mmHg, -2 mm: 18.8±0.5 mmHg, -6 mm: 22.8±0.3 mmHg) and 19 mm Mitroflow (+2 mm: 24.1±0.2 mmHg, -2 mm: 31.5±0.3 mmHg, -6 mm: 25.6±0.2 mmHg) valves had elevated mean gradients. In larger sized surgical valves (≥23 mm) the regurgitant fraction was higher at low implantation depths. Pinwheeling was significantly worse in the smaller sized (≤21 mm) surgical valves and also at low (<-2 mm) implantation depth.
The 23 mm ALLEGRA valve had favourable (<20 mmHg) gradients in all surgical valves sized ≥21 mm, even when the THV was implanted low. In 19 mm sized Mitroflow and Epic valves, gradients were elevated (>20 mmHg). While there was no major difference in mean transvalvular gradients, leaflet pinwheeling was worse at lower implantation depths.
我们旨在评估在三种不同外科瓣膜设计中,使用 ALLEGRA 经导管心脏瓣膜(THV)进行瓣膜内(VIV)介入时,植入深度对水动力功能的影响。
对于 19mm、21mm、23mm 和 25mm Epic、Mitroflow 和 Magna Ease 生物瓣,使用 23mm ALLEGRA THV 进行了多种植入深度(+2mm、-2mm 和-6mm)的测试。在每个植入深度都进行了多模态成像和水动力评估。在 Mitroflow 和 Epic 瓣膜尺寸≥21mm 以及所有 Magna Ease 瓣膜尺寸中,23mm ALLEGRA 瓣膜的梯度均<20mmHg。在较低的植入深度时,梯度没有明显增加。19mm Epic(+2mm:20.1±0.6mmHg,-2mm:18.8±0.5mmHg,-6mm:22.8±0.3mmHg)和 19mm Mitroflow(+2mm:24.1±0.2mmHg,-2mm:31.5±0.3mmHg,-6mm:25.6±0.2mmHg)瓣膜的平均梯度升高。在较大尺寸的外科瓣膜(≥23mm)中,在较低的植入深度时,反流分数更高。在较小尺寸(≤21mm)的外科瓣膜中,以及在较低的(<-2mm)植入深度时,叶片风车效应明显更差。
在所有尺寸≥21mm 的外科瓣膜中,即使 THV 植入较低,23mm ALLEGRA 瓣膜也具有有利的(<20mmHg)梯度。在 19mm 尺寸的 Mitroflow 和 Epic 瓣膜中,梯度升高(>20mmHg)。虽然跨瓣梯度的平均值没有明显差异,但叶片风车效应在较低的植入深度时更差。