An analytical method based on high-performance liquid chromatography-tandem mass spectrometry (HPLC-MS/MS) was developed for detecting chlorate and perchlorate residues in milk power. Chlorate and perchlorate in milk power were extracted using a 0.1% (v/v) formic acid-acetonitrile solution. The extract was centrifuged at 10000 r/min for 10 min, and the supernatant was cleaned up on a PRiME HLB column. Separation of chlorate and perchlorate was performed on an ion-exchange column (Thermo Scientific Acclaim TRINITY P1, 50 mm×2.1 mm, 3 μm) by gradient elution using acetonitrile and 20 mmol/L ammonium acetate solution as the mobile phase. The analytes were identified by MS/MS. Quantification was achieved using internal standards. Chlorate and perchlorate demonstrated good linearity in the ranges of 2.0-40.0 and 1.0-20.0 μg/L, respectively, with correlation coefficients () greater than 0.999. The limits of quantification (LOQs) of chlorate and perchlorate were found to be 15.0 and 7.5 μg/kg, respectively. The recoveries of chlorate and perchlorate ranged from 89.24% to 107.85% at the three spike levels of 30.0, 60.0, and 120.0, and 15.0, 30.0, and 60.0 μg/kg, respectively, with relative standard deviations (RSDs) ranging from 3.15% to 10.42% (=6). This method is convenient, rapid, accurate, and efficient, thus demonstrating its suitability for use in the determination of chlorate and perchlorate in milk power.
建立了一种基于高效液相色谱-串联质谱(HPLC-MS/MS)的分析方法,用于检测奶粉中氯酸盐和高氯酸盐残留。采用0.1%(v/v)甲酸-乙腈溶液提取奶粉中的氯酸盐和高氯酸盐。提取物在10000 r/min下离心10 min,上清液通过PRiME HLB柱净化。采用乙腈和20 mmol/L乙酸铵溶液作为流动相,通过梯度洗脱在离子交换柱(Thermo Scientific Acclaim TRINITY P1,50 mm×2.1 mm,3 μm)上分离氯酸盐和高氯酸盐。通过MS/MS鉴定分析物。使用内标进行定量。氯酸盐和高氯酸盐在2.0 - 40.0和1.0 - 20.0 μg/L范围内分别表现出良好的线性,相关系数()大于0.999。氯酸盐和高氯酸盐的定量限(LOQ)分别为15.0和7.5 μg/kg。在30.0、60.0和120.0 μg/kg以及15.0、30.0和60.0 μg/kg这三个加标水平下,氯酸盐和高氯酸盐的回收率分别为89.24%至107.85%,相对标准偏差(RSD)为3.15%至10.42%(=6)。该方法简便、快速、准确、高效,适用于奶粉中氯酸盐和高氯酸盐的测定。