De Giacinto Chiara, D'Aloisio Rossella, Bova Alessandro, Candian Tommaso, Perrotta Alberto Armando, Tognetto Daniele
Eye Clinic, Department of Medicine, Surgery and Health Sciences, University of Trieste, Trieste 34129, Italy.
J Ophthalmol. 2019 Sep 25;2019:5986895. doi: 10.1155/2019/5986895. eCollection 2019.
The aim of this retrospective cohort study was to evaluate intraocular pressure (IOP) changes during femtosecond laser-assisted cataract surgery (FLACS) using two different patient interface systems.
116 eyes of 116 patients scheduled for cataract surgery were divided into 2 groups: group 1 (61 eyes) and group 2 (55 eyes) underwent FLACS using Catalys Laser with fluid interface (liquid optics interface, LOI) and LenSx Laser with curved interface and soft contact lens (SoftFit), respectively. IOP was assessed using a portable rebound tonometer (Icare®) preoperatively, after docking, immediately after surgery, at one and seven days postoperatively.
In group 1, the mean IOP (±SD) was 14.1 ± 0.4 mmHg before surgery, 33.2 ± 1.1 mmHg after docking, and 21.4 ± 0.9 mmHg immediately after surgery. In group 2, the mean IOP was 13.8 ± 0.4 mmHg before surgery, 24.2 ± 1.4 mmHg after docking, and 20.2 ± 1.2 mmHg immediately after surgery. After the docking procedure, a statistically significant increase in IOP from the baseline was found in both groups ( < 0.001). Moreover, no statistically significant difference in IOP measured at 1 and 7 days postoperatively was observed compared with the preoperative values ( > 0.05) using both laser platforms. No intraoperative and postoperative complications were observed.
FLACS suction phase resulted in a transient increase of IOP in both groups, especially with the LOI system, and it is probably related to the greater pressure of a suction ring and suction generated through the vacuum, independently from the effect of femtosecond laser itself.
本回顾性队列研究的目的是评估使用两种不同患者接口系统的飞秒激光辅助白内障手术(FLACS)期间的眼压(IOP)变化。
116例计划进行白内障手术的患者的116只眼被分为2组:第1组(61只眼)和第2组(55只眼)分别使用具有流体接口(液体光学接口,LOI)的Catalys激光和具有曲面接口及软性接触镜(SoftFit)的LenSx激光进行FLACS。术前、对接后、手术后即刻、术后1天和7天使用便携式回弹眼压计(Icare®)评估眼压。
在第1组中,术前平均眼压(±标准差)为14.1±0.4 mmHg,对接后为33.2±1.1 mmHg,手术后即刻为21.4±0.9 mmHg。在第2组中,术前平均眼压为13.8±0.4 mmHg,对接后为24.2±1.4 mmHg,手术后即刻为20.2±1.2 mmHg。对接程序后,两组眼压均较基线有统计学显著升高(<0.001)。此外,使用两种激光平台时,术后1天和7天测量的眼压与术前值相比无统计学显著差异(>0.05)。未观察到术中及术后并发症。
FLACS吸引阶段导致两组眼压短暂升高,尤其是使用LOI系统时,这可能与吸引环的更大压力以及通过真空产生的吸力有关,与飞秒激光本身的作用无关。