Measuring diaphragm movement and respiratory frequency using a novel ultrasound device in healthy volunteers.

机构信息

Department of Pulmonary Medicine, Oslo University Hospital, Kirkeveien 166, 0450, Oslo, Norway.

Oslo Centre for Biostatistics and Epidemiology, Research Support Services, Oslo University Hospital, Oslo, Norway.

出版信息

J Ultrasound. 2021 Mar;24(1):15-22. doi: 10.1007/s40477-019-00412-2. Epub 2019 Nov 6.

Abstract

PURPOSE

To evaluate the ability of a novel ultrasound (US) device, DiaMon, to monitor diaphragm movement via its proxy liver movement, and compare it with the respired flow measured with a flowmeter, in awake and healthy volunteers. We wanted to (1) establish the optimal anatomical position for attaching the DiaMon device to the abdominal wall, and (2) evaluate the accuracy of continuous monitoring of respiratory frequency.

METHODS

Thirty healthy subjects were recruited. The DiaMon probe was applied subcostally in four different positions with the subjects in five different postures. The subjects breathed tidal volumes into a spirometer for 30-60 s with the DiaMon recording simultaneously.

RESULTS

The device detected a readable signal in 83-100% of the position/posture-combinations. The technical correlation between the two signals was highest in the anterior axillary-supine position (mean ± SD: 0.95 ± 0.03), followed by paramidline-supine (0.90 ± 0.09) and midclavicular-supine (0.89 ± 0.12). The frequency measurements yielded a mean difference of 0.03 (95% limits of agreement - 0.11, 0.16) breaths per minute in the anterior axillary-supine position.

CONCLUSION

The DiaMon device is able to detect liver movement in most subjects, and it measures breathing frequency accurately.

摘要

目的

评估新型超声(US)设备 DiaMon 通过其代理肝运动监测膈运动的能力,并将其与流量计测量的呼吸流量进行比较,用于清醒和健康志愿者。我们希望(1)确定将 DiaMon 设备附着到腹壁的最佳解剖位置,以及(2)评估连续监测呼吸频率的准确性。

方法

招募了 30 名健康受试者。将 DiaMon 探头以四种不同位置应用于肋缘下,受试者处于五种不同姿势。受试者通过呼吸计呼吸潮气量 30-60 秒,同时 DiaMon 进行记录。

结果

在 83-100%的位置/姿势组合中,该设备检测到可读信号。两种信号之间的技术相关性在前腋位仰卧位最高(平均 ± 标准差:0.95 ± 0.03),其次是中线仰卧位(0.90 ± 0.09)和锁骨中线仰卧位(0.89 ± 0.12)。在前腋位仰卧位,频率测量的平均差异为 0.03(95%一致性界限-0.11,0.16)每分钟呼吸次数。

结论

DiaMon 设备能够在大多数受试者中检测到肝运动,并且能够准确测量呼吸频率。

相似文献

1
Measuring diaphragm movement and respiratory frequency using a novel ultrasound device in healthy volunteers.
J Ultrasound. 2021 Mar;24(1):15-22. doi: 10.1007/s40477-019-00412-2. Epub 2019 Nov 6.
2
Comparison of Diaphragm Thickness Measurements Among Postures Via Ultrasound Imaging.
PM R. 2017 Jan;9(1):21-25. doi: 10.1016/j.pmrj.2016.06.001. Epub 2016 Jun 11.
3
Position and motion of the human diaphragm during anesthesia-paralysis.
Anesthesiology. 1989 Jun;70(6):891-8. doi: 10.1097/00000542-198906000-00002.
6
Assessing the impact of posture on diaphragm morphology and function using an open upright MRI system-A pilot study.
Eur J Radiol. 2020 Sep;130:109196. doi: 10.1016/j.ejrad.2020.109196. Epub 2020 Jul 25.
8
The effect of posture on asynchronous chest wall movement in COPD.
J Appl Physiol (1985). 2013 Apr;114(8):1066-75. doi: 10.1152/japplphysiol.00414.2012. Epub 2013 Feb 14.
10

引用本文的文献

1
Breathless Strength: Ultrasonographic Insights into Expiratory Muscle Dysfunction in Spinal Cord Injury.
Medicina (Kaunas). 2025 May 15;61(5):897. doi: 10.3390/medicina61050897.
2
Ultrasonographic reliability and repeatability of simultaneous bilateral assessment of diaphragm muscle thickness during normal breathing.
Quant Imaging Med Surg. 2023 Oct 1;13(10):6656-6667. doi: 10.21037/qims-23-329. Epub 2023 Aug 17.
4
Evaluation of diaphragm thickening by diaphragm ultrasonography: a reproducibility and a repeatability study.
J Ultrasound. 2021 Dec;24(4):411-416. doi: 10.1007/s40477-020-00462-x. Epub 2020 May 1.

本文引用的文献

2
Improving detection of patient deterioration in the general hospital ward environment.
Eur J Anaesthesiol. 2018 May;35(5):325-333. doi: 10.1097/EJA.0000000000000798.
7
Ultrasound in obstructive lung diseases: the effect of airway obstruction on diaphragm kinetics. A short pictorial essay.
J Ultrasound. 2014 Aug 1;18(4):379-84. doi: 10.1007/s40477-014-0122-5. eCollection 2015 Dec.
8
Design and Evaluation of an Intelligent Remote Tidal Volume Variability Monitoring System in E-Health Applications.
IEEE J Biomed Health Inform. 2015 Sep;19(5):1532-48. doi: 10.1109/JBHI.2015.2445783. Epub 2015 Jun 16.
9
The accuracy of respiratory rate assessment by doctors in a London teaching hospital: a cross-sectional study.
J Clin Monit Comput. 2015 Aug;29(4):455-60. doi: 10.1007/s10877-014-9621-3. Epub 2014 Oct 2.
10
Is diaphragm motion a good surrogate for liver tumor motion?
Int J Radiat Oncol Biol Phys. 2014 Nov 15;90(4):952-8. doi: 10.1016/j.ijrobp.2014.07.028. Epub 2014 Sep 12.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验