J Ultrasound. 2021 Mar;24(1):15-22. doi: 10.1007/s40477-019-00412-2. Epub 2019 Nov 6.
To evaluate the ability of a novel ultrasound (US) device, DiaMon, to monitor diaphragm movement via its proxy liver movement, and compare it with the respired flow measured with a flowmeter, in awake and healthy volunteers. We wanted to (1) establish the optimal anatomical position for attaching the DiaMon device to the abdominal wall, and (2) evaluate the accuracy of continuous monitoring of respiratory frequency.
Thirty healthy subjects were recruited. The DiaMon probe was applied subcostally in four different positions with the subjects in five different postures. The subjects breathed tidal volumes into a spirometer for 30-60 s with the DiaMon recording simultaneously.
The device detected a readable signal in 83-100% of the position/posture-combinations. The technical correlation between the two signals was highest in the anterior axillary-supine position (mean ± SD: 0.95 ± 0.03), followed by paramidline-supine (0.90 ± 0.09) and midclavicular-supine (0.89 ± 0.12). The frequency measurements yielded a mean difference of 0.03 (95% limits of agreement - 0.11, 0.16) breaths per minute in the anterior axillary-supine position.
The DiaMon device is able to detect liver movement in most subjects, and it measures breathing frequency accurately.
评估新型超声(US)设备 DiaMon 通过其代理肝运动监测膈运动的能力,并将其与流量计测量的呼吸流量进行比较,用于清醒和健康志愿者。我们希望(1)确定将 DiaMon 设备附着到腹壁的最佳解剖位置,以及(2)评估连续监测呼吸频率的准确性。
招募了 30 名健康受试者。将 DiaMon 探头以四种不同位置应用于肋缘下,受试者处于五种不同姿势。受试者通过呼吸计呼吸潮气量 30-60 秒,同时 DiaMon 进行记录。
在 83-100%的位置/姿势组合中,该设备检测到可读信号。两种信号之间的技术相关性在前腋位仰卧位最高(平均 ± 标准差:0.95 ± 0.03),其次是中线仰卧位(0.90 ± 0.09)和锁骨中线仰卧位(0.89 ± 0.12)。在前腋位仰卧位,频率测量的平均差异为 0.03(95%一致性界限-0.11,0.16)每分钟呼吸次数。
DiaMon 设备能够在大多数受试者中检测到肝运动,并且能够准确测量呼吸频率。