心脏植入式电子设备患者的心脏磁共振:挑战与解决方案。

Cardiac Magnetic Resonance in Patients With Cardiac Implantable Electronic Devices: Challenges and Solutions.

机构信息

Department of Radiology, Division of Cardiothoracic Imaging, UT Southwestern Medical Center, Dallas, TX.

Cardiovascular Imaging Laboratory, Cleveland Clinic Foundation, Cleveland, OH.

出版信息

J Thorac Imaging. 2020 Jan;35(1):W1-W17. doi: 10.1097/RTI.0000000000000462.

Abstract

Until recently, cardiac implantable electronic devices (CIEDs) were an absolute contraindication to magnetic resonance imaging (MRI), due to concerns about their adverse interaction in the MRI environment. The increasing clinical need to perform MRI examinations in these patients was an impetus to the development of MR-Conditional CIEDs. Secure performance of MRI in these patients requires scanning under specified MR conditions as well as operating the device in MR-scanning mode. This requires robust institutional protocols and a well-trained multidisciplinary team of radiologists, cardiologists, device applications specialists, physicists, nurses, and MRI technologists. MRI can also be performed in patients with non-MRI Conditional or "legacy" CIEDs by following safety precautions and continuous monitoring. Cardiac magnetic resonance (CMR) is additionally challenging due to expected susceptibility artifacts generated by the CIEDs, which are either near or in the heart. As the most common indication for CMR in these patients is the evaluation of myocardial scar/fibrosis, acquiring a high-quality late gadolinium enhancement image is of the utmost importance. This sequence is hampered by artifactual high signal due to inadequate myocardial nulling. Several solutions are available to reduce these artifacts, including reducing inhomogeneity, technical adjustments, and use of sequences that are more resilient to artifacts. In this article, we review the precautions for CMR in patients with CIEDs, provide guidelines for secure performance of CMR in these patients, and discuss techniques for obtaining high quality CMR images with minimized artifacts.

摘要

直到最近,心脏植入式电子设备 (CIEDs) 由于在 MRI 环境中存在不良相互作用的担忧,一直被视为绝对禁忌,不能进行磁共振成像 (MRI)。在这些患者中进行 MRI 检查的临床需求不断增加,推动了 MRI 条件下 CIED 的发展。要在这些患者中安全地进行 MRI,需要在特定的 MRI 条件下进行扫描,并以 MRI 扫描模式操作设备。这需要强大的机构协议和训练有素的多学科团队,包括放射科医生、心脏病专家、设备应用专家、物理学家、护士和 MRI 技术人员。对于非 MRI 条件或“传统”CIED 患者,通过遵循安全预防措施和持续监测,也可以进行 MRI。由于 CIED 产生的预期磁化率伪影,心脏磁共振 (CMR) 也具有挑战性,这些伪影要么靠近心脏,要么就在心脏内。由于这些患者进行 CMR 的最常见适应症是评估心肌瘢痕/纤维化,因此获得高质量的钆延迟增强图像至关重要。由于心肌未完全消除,该序列会因伪影导致高信号。有几种解决方案可以减少这些伪影,包括减少不均匀性、技术调整以及使用对伪影更具弹性的序列。在本文中,我们回顾了 CIED 患者进行 CMR 的注意事项,提供了在这些患者中安全进行 CMR 的指南,并讨论了获得高质量 CMR 图像和最小化伪影的技术。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验