We developed an effective hemostatic method using Hydrofit and a hemostatic gelatin sponge (Spongel). We evaluated the hemostatic effect in comparison to the conventional silicone sheet method.
A simulated circuit was created using the pump of a Nipro ventricular assist system and a prosthetic graft. A hole was made in the graft by a needle and three hemostatic methods were applied: the silicone sheet method (SS) using Hydrofit and a silicone sheet, the bread and butter method (BB) using Hydrofit and a gelatin sponge instead of a silicone sheet, and French toast method (FT) using Hydrofit and a gelatin sponge over which water was poured before compression. The amount of leakage before and after the application each of the methods was measured according to the compression time.
In the 60 s compression, the amount of leakage after SS, BB, and FT was 0.4 ± 0.8, 0.2 ± 0.6, and 0 ± 0.0 ml, respectively, and FT showed no leakage. In the 30 s compression, the amount of leakage after SS, BB, and FT was 14.2 ± 27.9, 1.0 ± 3.2, and 7.8 ± 22.6 ml, respectively, and did not differ to a statistically significant extent.
The method of combining Hydrofit and Spongel could obtain reliable hemostasis in 60 s.
我们开发了一种使用 Hydrofit 和止血明胶海绵(Spongel)的有效止血方法。我们将其止血效果与传统的硅胶片方法进行了比较。
使用尼普洛心室辅助系统的泵和人造移植物创建了一个模拟回路。在移植物上用针打一个孔,并应用三种止血方法:使用 Hydrofit 和硅胶片的硅胶片法(SS)、使用 Hydrofit 和明胶海绵代替硅胶片的面包黄油法(BB)以及在压缩前先用水浇湿明胶海绵的法式吐司法(FT)。根据压缩时间测量每种方法应用前后的泄漏量。
在 60 秒的压缩中,SS、BB 和 FT 之后的泄漏量分别为 0.4±0.8、0.2±0.6 和 0.0±0.0ml,FT 没有泄漏。在 30 秒的压缩中,SS、BB 和 FT 之后的泄漏量分别为 14.2±27.9、1.0±3.2 和 7.8±22.6ml,没有统计学意义上的显著差异。
将 Hydrofit 和 Spongel 结合的方法可以在 60 秒内获得可靠的止血效果。