Suppr超能文献

使用新型、多功能且价格低廉的导管进行实时输尿管识别。

Real-time ureteral identification with novel, versatile, and inexpensive catheter.

作者信息

Ushimaru Yuki, Ohigawa Atsushi, Yamashita Kotaro, Saito Takuro, Tanaka Koji, Makino Tomoki, Takahashi Tsuyoshi, Kurokawa Yukinori, Yamasaki Makoto, Mori Masaki, Doki Yuichiro, Nakajima Kiyokazu

机构信息

Department of Next Generation Endoscopic Intervention (Project ENGINE), Center of Medical Innovation and Translational Research, Osaka University Graduate School of Medicine, Suite 0912, 2-2, Yamadaoka, Suita, Osaka, 565-0871, Japan.

Department of Gastroenterological Surgery, Osaka University Graduate School of Medicine, Osaka, Japan.

出版信息

Surg Endosc. 2020 Aug;34(8):3669-3678. doi: 10.1007/s00464-019-07261-4. Epub 2020 Jan 6.

Abstract

BACKGROUND

Although ureteral catheters and ureteral fluorescence methods have been investigated to avoid ureteral injury, they have not been standardized for procedural complexity and safety to the living body. A near-infrared (NIR) fluorescence ureteral catheter made of fluorescent resin was jointly developed for non-invasive detection of ureters. The aims of this study were to (1) evaluate its bench-top performance and (2) assert its safety and potential usefulness in a series of animal models.

METHODS

[Bench-top study]: Confirmed stimulation of NIR fluorescence catheter by NIR light was investigated with the use of a laparoscopic fluorescence imaging system. Additionally, the influence of imaging distance and shielding objects, such as 1.5-mm sliced pig loin with multiple sheets, was also evaluated. [Performance study]: The safety and potential usefulness of fluorescence catheter was evaluated in five pigs. Non-fluorescence and fluorescence ureteral catheters were alternatively placed in the animals' left and right ureters. The ImageJ software was used in all experiments to quantify fluorescence signal and the signal-to-background ratio.

RESULTS

[Bench-top study]: The fluorescence ureteral catheter was successfully identified at all distances. Its fluorescence decreased in inverse proportion to distance and to the intervening shield thickness (p  < 0.01). When shields were present, catheter position could not be recognized with non-fluorescence catheters, but with fluorescence catheters they could be recognized. [Performance study]: Fluorescence catheter's ability to fluoresce at all distances was confirmed (p  < 0.01). No individual differences (p = 0.21) or left/right ureter differences (p  = 0.79) were observed. The fluorescence of the fluorescence catheter decreased in inverse proportion to distance (p  < 0.01).

CONCLUSIONS

The new fluorescence ureter catheter investigated shows promising performance in providing ureteral identification with high specificity during laparoscopic surgery.

摘要

背景

尽管已经对输尿管导管和输尿管荧光方法进行了研究以避免输尿管损伤,但由于操作复杂性和对活体的安全性,它们尚未标准化。一种由荧光树脂制成的近红外(NIR)荧光输尿管导管被联合开发用于输尿管的无创检测。本研究的目的是:(1)评估其在实验台上的性能;(2)在一系列动物模型中确定其安全性和潜在实用性。

方法

[实验台研究]:使用腹腔镜荧光成像系统研究近红外光对近红外荧光导管的确认刺激。此外,还评估了成像距离和屏蔽物体(如多层1.5毫米厚的猪里脊肉切片)的影响。[性能研究]:在五只猪中评估荧光导管的安全性和潜在实用性。将非荧光和荧光输尿管导管交替放置在动物的左右输尿管中。在所有实验中使用ImageJ软件量化荧光信号和信号背景比。

结果

[实验台研究]:在所有距离都成功识别出荧光输尿管导管。其荧光与距离和中间屏蔽厚度成反比下降(p < 0.01)。当存在屏蔽物时,非荧光导管无法识别导管位置,但荧光导管可以识别。[性能研究]:确认了荧光导管在所有距离都能发出荧光(p < 0.01)。未观察到个体差异(p = 0.21)或左右输尿管差异(p = 0.79)。荧光导管的荧光与距离成反比下降(p < 0.01)。

结论

所研究的新型荧光输尿管导管在腹腔镜手术期间提供高特异性输尿管识别方面显示出有前景的性能。